This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.
Products
The products listed here are for overseas markets, not for Japan domestic.
3-in-1 safety tester integrating AC Withstand Voltage, DC Withstand Voltage, and Insulation Resistance (IR) measurement functions.
The 7640/7641 Hipot Tester is a 3-in-1 safety tester integrating AC Withstand Voltage, DC Withstand Voltage, and Insulation Resistance (IR) measurement functions. It provides AC hipot output up to 5kV with leakage current up to 30mA, and DC hipot output up to 6kV with leakage current up to 10mA. The insulation resistance test supports output voltage up to 6kV, with a measurement range from 200kΩ~100GΩ. For production line applications, the system supports multi-channel configurations: the 7641 offers up to 8 test channels per unit and can be expanded via 7508A / 7516A scanning boxes to a maximum of 72 channels. In addition, the instrument supports PC connectivity software and USB storage for test data.
Hipot Tester
Model:
7640 (Single Channel)
Maker:
MICROTEST Corporation
Features
- AC Withstand Voltage up to 5000V
- DC Withstand Voltage up to 6000V
- Insulation Resistance up to 100GΩ
- Supports up to 72-Channel Test (7508A / 7516A)
- Programmable Ramp / Dwell / Fall Time (0.1s~99.9s)
- Arc Detection Function
- True RMS measurement circui
- Safety Interlock Protection
- PC Connectivity and Remote Control Support
Specifications
| Model | 7640 |
| Channel | 1 |
| AC Hipot | |
| Output Voltage | 10V~5000V |
| Voltage Resolution | 1V |
| Voltage Accuracy | ±(2% of setting +5V) no load |
| Output Frequency | 50/60Hz |
| Rated Current (Max.) | 30mA |
| Leakage Current | 0~31mA |
| Lower Leakage Current Resolution | 0.001mA |
| Lower Leakage Current Accuracy | ±(1% of reading +5uA) |
| Measuring Time | 0.1~999s, 0 (Continuous) |
| Measuring Time Resolution | 0.1s |
| Ramp, Dwell, Fall Time | 0.1~99.9 sec |
| Arcing Detection | Level 0 for Off / Adjustable levels 1~9 (Level 1 is the most sensitive) |
| DC Hipot | |
| Output Voltage | 10V~6000V |
| Voltage Resolution | 1V |
| Voltage Accuracy | ±(2% of setting +5V) |
| Rated Current (Max.) | 10mA |
| Leakage Current | 0~11mA |
| Lower Leakage Current Resolution | 0.001mA |
| Lower Leakage Current Accuracy | ±(1% of reading +5uA) |
| Measuring Time | 0.1~999s, 0 (Continuous) |
| Measuring Time Resolution | 0.1s |
| Ramp, Dwell, Fall Time | 0.1~99.9 sec |
| Arcing Detection | Level 0 for Off / Adjustable levels 1~9 (Level 1 is the most sensitive) |
| Insulation Resistance | |
| Output Voltage | 10V~6000V |
| Voltage Resolution | 1V |
| Voltage Accuracy | ±(2% of setting+5V) |
| Rated Current (Max.) | 1mA |
| IR Upper Threshold | 200k~100GΩ (Current measurement range 5nA ≦ i ≦ 1mA) |
| IR Resolution | 0.001MΩ |
| IR Accuracy | 50V { 200k~500MΩ, ±(5% of reading+30kΩ)/ 500M~1GΩ, ±(10% of reading+20MΩ)/ 1G~10GΩ, ±(25% of reading+100MΩ)} |
| 100V { 200k~200MΩ, ±(2% of reading+30kΩ)/ 200M~1GΩ, ±(5% of reading+20MΩ)/ 1G~20GΩ, ±(25% of reading+100MΩ)} | |
| 500V { 500k~500MΩ, ±(2% of reading+30kΩ)/ 500M~10GΩ, ±(5% of reading+20MΩ)/ 10G~50GΩ, ±(20% of reading+100MΩ)} | |
| ≥ 1000V { 1M~1GΩ, ±(2% of reading+30kΩ)/ 1G~10GΩ, ±(5% of reading+20MΩ)/ 10G~100GΩ, ±(20% of reading+100MΩ)} | |
| Measuring Time | 0.1~999s, 0 (Continuous) |
| Measuring Time Resolution | 0.1s |
| Ramp, Dwell, Fall Time | 0.1~99.9 sec |
| Remote Control | Test、About |
| Remote Output Signal | Pass、Fail、HV output、Testing |
| Safety Switch | When testing, you need to short-circuit the INTER LOCK on the rear of the instrument to output the test voltage |
| Built-in Storage | 7640(64 sets), 7641(32 sets) |
| USB Host Storage | Storage setting files and update firmware; the number of saved data sets depends on the USB capacity |
| List Mode | Each group contains up to 8 (7640)/ 16 (7641) steps |
| Interface | RS-232、Remote、USB Host/Device、EXT.I/O |
| Power Supply | Voltage:100Vac~240Vac/ Frequency:50/60Hz ±5% |
| Power Consumption | 350VA |
| Display | 480*272, 4.3″ TFT LCD |
| Environment | Temperature:5℃~40℃, Humidity:20~80%RH |
| Dimension (W*H*D) | 340×88×369.6mm(7640)/ 340×88×429.6mm(7641) |
| Weight | 12Kg(7640)/ 14Kg(7641) |