This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.
Products
The products listed here are for overseas markets, not for Japan domestic.
Supports AC withstand voltage / DC withstand voltage / insulation impedance measurement functions, capable of storing measurement data via USB, and supports PC connection software.
MICROTEST 7631 is a 3-in-1 test instrument that supports AC withstand voltage, DC withstand voltage, and insulation resistance testing functions. It has a maximum AC withstand voltage of 5000V with a leakage current of 30mA, a maximum DC withstand voltage of 6000V with a leakage current of 10mA, and a maximum insulation resistance of 12000MΩ. It provides 8 sets of testing channels and can be optionally equipped with the 7508/7516 multi-channel scanning box, allowing up to 72 sets of testing channels. It supports PC connection software for remote programming and storage of test data. It can also be linked to a barcode scanner for barcode scanning, facilitating production traceability management.
Hipot Tester
Model:
7631 (8 Channel)
Maker:
MICROTEST Corporation
Features
- 8 Channels
- Hi pot ACW 5KV/30mA
- Hi pot DCW 6KV/10mA
- Insulation 1KV/12000MΩ
- Arcing detection
- True RMS measurement circuit
- Measuring Time (0.1 ~ 999s)
- Adjustable ramp time (0.1 ~ 10s)
- Safety Switch(INTER LOCK)
- Option PC Link Software
- 7508/7516 (8/16 channels) expansion box
- Support AC 5000V, DC 6000V, Insulation Resistance 12GΩ
Videos
Specifications
| Model | 7631 | |
| Test Channel | 8 | |
| AC Output Voltage | 5000V | |
| DC Output Voltage | 6000V | |
| Insulation Resistance | 12000MΩ | |
| Expansion box | 8 Chnnels | 16 Channels |
| 7508 | 7516 | |
| AC Hipot | ||
| Output Voltage | 10V-5000V | |
| Voltage Resolution | 1V | |
| Voltage Accuracy | ±(3% of setting+5V) | |
| Rated Current (Max.) | 30mA | |
| Upper Leakage Current | 0.001-31mA | |
| Leakage Current Resolution | 0.001mA | |
| Leakage Current Accuracy | ±(3% of reading +10uA) | |
| Arcing Detection | Detect gears from 0-20 (1 is the most sensitive) | |
| DC Hipot | ||
| Output Voltage | 10V-6000V | |
| Voltage Resolution | 1V | |
| Voltage Accuracy | ±(3% of setting+5V) | |
| Rated Current (Max.) | 10mA | |
| Upper Leakage Current | 0.001-11mA | |
| Leakage Current Resolution | ±(3% of reading +10uA) | |
| Leakage Current Accuracy | 0.001mA | |
| Arcing Detection | Detect gears from 0-10 (1 is the most sensitive) | |
| Insulation Resistance | ||
| Output Voltage | 10V-1000Vdc | |
| Voltage Resolution | 1V | |
| Voltage Accuracy | ±(3% of setting+5V) | |
| Insulation Resistance Range | 1-12000M? | |
| IR Resolution | 1M? | |
| IR Accuracy | 100-500V, 1-1000MΩ (±5% of reading+1MΩ)
500-1000V, 2-12000MΩ (±5% of reading +10MΩ) |
|
| Measuring Time | 0.1-999s | |
| Ramp Time | 0.1-10s | |
| Measuring Time Resolution | 0.1s | |
SYSTEM
| PLC Remote Control | Test、About |
| PLC Output Signal | PASS、FAIL、HV Output、Testing |
| Safety Switch | When testing, you need to short-circuit the INTER LOCK on the rear of the instrument to output the test voltage |
| List Mode | Each group contains up to 32 steps |
| Built-in Storage | 15sets setting files |
| USB Host Storage | Storage setting files and update firmware |
| Power Supply | Voltage:90Vac-264Vac |
| Frequency:50/60Hz ±5% | |
| Power consumption | 300VA |
| Environment | Temperature:5℃-40℃ |
| Humidity:0-80%RH | |
| Dimension | 435x145x500 mm (W*H*D) |
| Weight | 15 Kg |
| Interface | RS-232、Remote、USB Host、USB Device、EXT. I/O |
| Display | 4.3″ TFT LCD (480*272 ) |