CONTENTS CONTENTS

Products

The products listed here are for overseas markets, not for Japan domestic.

MICROTEST 3-in-1 Transformer Testing System (Low Voltage+Safety+Impulse)

20 Channels Transformer Tester

The MICROTEST 2-in-1 Transformer Integrated Test System is a comprehensive solution that combines low-voltage electrical characteristics testing and safety compliance testing. It offers 20 testing channels and integrates low-voltage testing (including inductance, leakage inductance, resistance, capacitance, etc.) with AC/DC withstand voltage insulation testing (up to 5000V/6000V). It ensures efficient and convenient testing for transformer products.

To measure the inductance value of a transformer, you can use an LCR meter, which offers selectable test frequencies including 200kHz, 500kHz, and 1MHz. It supports software that allows connection to a PC, enabling the storage of test data.

3-in-1 Transformer Testing System (Low Voltage+Safety+Impulse)

Model:

6265+7605+7721, 6266+7605+7721, 6267+7605+7721

Maker:

MICROTEST Corporation

Features

  • Low Voltage Electrical Test
  • Hi pot/ Insulation Test
  • Impulse Test (Layer Short Circuit)
  • 20 Channels
  • Test Frequency (200k/500k/1MHz)
  • AC Voltage Output 5000V
  • DC Voltage Output 6000V
  • Insulation Resistance 12000MΩ

Videos

Specifications

Model 3-in-1 (626X+7605+7721)
Test Channel

20

Low Voltage Electrical Test
Model 6265 6266 6267
Frequency 10Hz-200kHz 10Hz-500kHz 10Hz-1MHz
Frequency Resolution 5 digits
Basic Accuracy ±0.1%
AC Drive Level 10mV-2Vrms
DC Drive Level 10mV-2V
Output Impedance 100Ω
Turn

Inductance or Voltage

Test Frequency:50Hz-200kHz

Test Mode Meter Mode / List Mode
Parameters Measurement Inductance (L)、Impedance (Z)、Capacitance (C)、Resistance (R)、Conductance (G)、Susceptance (B)、Admittance (Y)、Alternating Current Resistance (ACR)、Quality Factor (Q)、θ、Direct Current Resistance (DCR)、Leakage Inductance、Turn Ratio、Balance、Short Circuit
L, LK 0.1nH ~ 9999.99H
C 0.00001pF ~ 999.99mF
Q,D 0.00001 ~ 99999
Z,X,R 0.00001Ω ~ 99.9999MΩ
Y 0.01nS ~ 99.9999S
θ -180°~ +180°
DCR 0.1mΩ ~ 99.999 MΩ
Turn 0.1 ~ 99999.9 turns
Pin-Short 12 pairs, between pin to pin
Hi pot/ Insulation Test
Items AC Hi pot DC Hi pot IR
Output Voltage 10V-5000V 10V-6000V 10V-1000V
Voltage Resolution 1V 1V 1V
Voltage Accuracy
Test Range 0.001-31mA 0.001-11mA 1-12000MΩ
AC Arcing Detection 0-20 0-10 –
Measuring time 0.1-999s
Ramp Time 0.1-10s
Time Resolution 0.1s
Impulse Test
Impulse Voltage 200V-5000V

Area

Comparison

When layer short happened, the loss of power on coil increase, the resonance damping coefficient increase, resonance amplitude decrease, the total area decrease. These are the basic parameters we check layer short.

By calculating and comparing the deference of area between golden sample and DUT.
Show by percentage
Analysis the power loss by this data.

Area

Differential

Add up the difference between normal wave and DUT wave call “Area differential”.
When layer short happened, the inductance reduce(similar the transformer). The wave phase change and the area differential also change. This will show “fail” on the instrument.
However, the parameter might cause deviation because of the deviation of inductance, resonance phase shift. (Silicon steel product such as motor and traditional transformer is not suitable)

By calculating and comparing the deference of area between golden sample and DUT. To determine the degree of waveform overlap.
Compare the inductance by this.

Waveform

Comparison

Set a acceptable wave range, if the DUT’s wave is in this range shows “pass” otherwise,”fail”.
This parameter can judge both amplitude and phase of resonance wave. This can increase the detection capability of layer short.

Corona In pulse test, the insulation defect will cause discharge and create corona. This function is able to count the times that corona happened base on the degree of deviation.
Detect the discharge phenomenon on the coil.
Flat

If the layer short happened, the waveform will tremble. Therefore, the instrument will quantize and compared it.

System

Model Transformer Tester Impulse Tester Hi pot Tester
6265/6266/6267 7721 7605

Built-in

storage

128 sets 200 sets –

Power

Supply

Voltage

98Vac-132Vac

or 195Vac-264Vac

Voltage

115/230Vac±15%

Voltage

98Vac-132Vac

or 195Vac-264Vac

Frequency:47-63Hz Frequency:50/60Hz±5% Frequency:47-63Hz

Power

consumption

600VA/set

Dimension

(W*H*D)

344x145x343 mm (W*H*D) 435x190x522 mm 435x145x500 mm(W*H*D)
Weight 9 Kg 14kg 15kg
Interface RS-232、Handler、LAN、USB Host、EXT. I/O RS-232、Remote、Printer RS-232、Remote、USB Host、USB Device、EXT.I/O
Display Color Screen, 7″ TFT
(800*480)
5.7″ TFT
(320*240)
Color Screen, 4.3″ TFT
(480*272)

Get Adobe' Reader'

To view the pdf files, the Adobe Reader from Adobe Systems is required.