KIKUSUI Electrical Safety Multi-Analyzer TOS9300 Series

The All-Rounder. Hipot, Insulation Resistance, Ground Bond, Leakage or Partial Dischage,this analyzer covers it all!

The TOS9300 series is a high-performance electrical safety analyzer that complies with a wide range of universal standards. Hipot, Insulation Resistance, Ground Bond, Leakage Current (touch current and protective conductor current) and partial discharge can all be tested. A total of 6 models are available for standard compliance tests for a wide variety of applications including R&D, quality assurance manufacturing lines and laboratory tests.

Electrical Safety Multi-Analyzer TOS9300 series

Model:

TOS9300, TOS9301, TOS9301PD, TOS9302, TOS9303, TOS9303LC

Maker:

KIKUSUI ELECTRONICS CORPORATION

Features

  • All-in-one safety tester model (TOS9303LC)
  • Insulation diagnosis available with partial discharge model (TOS9301PD)
  • New amplifier type allows for 40 A AC/DC ground bond testing (Ground bond tester models)
  • Electrical breakdown inspection setting available
  • AC5 kV/100 mA, DC7.2 kV/100 W Hipot test
  • Touch current/protective conductor current/leakage current test (TOS9303LC)
  • LAN/USB/RS232C standard digital interface
  • Easy to read LCD display for real time monitoring during tests. All measurement values and standard outlines displayed in each test
  • High voltage scanner capable of output distribution both standalone and when connected with existing withstanding voltage/insulation resistance testing equipment models [TOS5300 series, etc.] (TOS9320)

Line up

Model Description Test Items Dimensions (Max) Weight
TOS9300 AC Hipot Tester with Insulation Resistance Test
  • ACW: 5 kV/100 mA (500 VA)
  • IR: 0.001 MΩ to 100.0 GΩ (DC -25 V to -1000 V)
430 (16.93″) (440 (17.32″)) W × 132 (5.2″) (155 (6.10″)) H × 370 (14.57″) (410 (16.14″)) D mm (inch) Approx. 17 kg (37.5 lbs)
TOS9301 AC/DC Hipot Tester with Insulation Resistance Test
  • ACW: 5 kV/100 mA (500 VA)
  • DCW: 5 kV/20 mA, 7.2 kV/13.9 mA (100 W)
  • IR: 0.001 MΩ to 100.0 GΩ (DC -25 V to -1000 V / DC +50 V to +7200 V)
430 (16.93″) (440 (17.32″)) W × 132 (5.2″) (155 (6.10″)) H × 370 (14.57″) (410 (16.14″)) D mm (inch) Approx. 18 kg (39.7 lbs)
TOS9301PD AC/DC Hipot Tester with Insulation Resistance and Partial Discharge Test
  • ACW: 5 kV/100 mA (500 VA)
  • DCW: 5 kV/20 mA, 7.2 kV/13.9 mA (100 W)
  • IR: 0.001 MΩ to 100.0 GΩ (DC -25 V to -1000 V / DC +50 V to +7200 V)
  • PD: 5 kV/50 mA (250 VA)
430 (16.93″) (440 (17.32″)) W × 132 (5.2″) (150 (5.9″)) H × 525 (20.67″) (565 (22.24″)) D mm (inch) Approx. 22 kg (48.5 lbs)
TOS9302 AC Hipot Tester with Ground Bond Test
  • ACW: 5 kV/100 mA (500 VA)
  • EC: 0.001 Ω to 0.600 Ω (3.0 A to 42.0 A)
430 (16.93″) (440 (17.32″)) W × 132 (5.2″) (155 (6.10″)) H × 500 (19.69″) (540 (21.26″)) D mm (inch) Approx. 20 kg (44.1 lbs)
TOS9303 AC/DC Hipot Tester with Insulation Resistance and Ground Bond Test
  • ACW: 5 kV/100 mA (500 VA)
  • DCW: 5 kV/20 mA, 7.2 kV/13.9 mA (100 W)
  • IR: 0.001 MΩ to 100.0 GΩ (DC -25 V to -1000 V / DC +50 V to +7200 V)
  • EC: 0.001 Ω to 0.600 Ω (3.0 A to 42.0 A)
430 (16.93″) (440 (17.32″)) W × 132 (5.2″) (155 (6.10″)) H × 500 (19.69″) (540 (21.26″)) D mm (inch) Approx. 21 kg (46.3 lbs)
TOS9303LC AC/DC Hipot Tester with Insulation Resistance, Ground Bond, and Leakage Current Test
  • ACW: 5 kV/100 mA (500 VA)
  • DCW: 5 kV/20 mA, 7.2 kV/13.9 mA (100 W)
  • IR: 0.001 MΩ to 100.0 GΩ (DC -25 V to -1000 V / DC +50 V to +7200 V)
  • EC: 0.001 Ω to 0.600 Ω (3.0 A to 42.0 A)
  • LC: 1 μA to 100 mA (rms)
430 (16.93″) (440 (17.32″)) W × 132 (5.2″) (155 (6.10″)) H × 500 (19.69″) (550 (21.65″)) D mm (inch) Approx. 22 kg (48.5 lbs)
Model AC Withstanding Voltage (AC Hipot) DC Withstanding Voltage (DC Hipot) Insulation Resistance Earth Continuity (Ground Bond) Leakage Current Partial Discharge
TOS9300 ● ●
TOS9301 ● ● ●
TOS9301PD ● ● ● ●
TOS9302 ● ●
TOS9303 ● ● ● ●
TOS9303LC ● ● ● ● ●
TOS9320 4 channel high voltage scanner with contact check function; can be used standalone.

Videos

Specifications

Item TOS9300 TOS9301 TOS9301PD TOS9302 TOS9303 TOS9303LC
Backup battery life 3 years (at 25 °C or 77 °F)
Environment Installation location Indoors, 2000 m or less
Spec guaranteed range Temperature 5 °C to 35 °C (41 °F to 95 °F)(18 °C to 28 °C for partial discharge tests)
Humidity 20 %rh to 80 %rh (20 %rh to 70 %rh for partial discharge tests)(no condensation)
Operating range Temperature 0 °C to 40 °C (32 °F to 104 °F)
Humidity 20 %rh to 80 %rh (no condensation)
Storage range Temperature -20 °C to 70 °C (-4 °F to 158 °F)
Humidity 90 %rh or less (no condensation)
Power supply Nominal voltage range (allowable voltage range) 100 Vac to 120 V, 200 V to 240 V (90 Vac to 132 V, 170 V to 250 V)
Power consumption No load(READY state) 100 VA or less
Rated load 800 VA max.
Allowable frequency range 47 Hz to 63 Hz
Insulation resistance (between AC LINE and chassis) 30 MΩ or more (500 Vdc)
Withstanding voltage (between AC LINE and chassis) 1500 Vac, 1 minute, 20 mA or less
Earth continuity 25 Aac, 0.1 Ω or less
Weight Approx. 17 kg Approx. 18 kg Approx. 22 kg Approx. 20 kg Approx. 21 kg Approx. 22 kg
Accessories Power cord (1 pc., *length: 2.5 m : The attached power cord varies depending on the shipment destination.), High-voltage test lead: TL31-TOS (1 pair), SIGNAL I/O plug (1 set), High-voltage warning sticker (1 pc.), Setup Guide (1 copy), CD-ROM (1 disc), Safety Information (1 copy), Heavy object warning label (1 pc., *Not included with the TOS9300), Test leads for earth continuity test: TL13-TOS (1 pair., *TOS9302, TOS9303,TOS9303LC only), [TOS9303LC only: Spare fuse (1 pc.), Test leads for leakage current test (2 red, 1 black), Flat probe (1 sheet)]
Electromagnetic compatibility *1 *2 Complies with the requirements of the following directive and standards. EMC Directive 2014/30/EU EN 61326-1 (Class A *3), EN 55011 (Class A *3, Group 1 *4), EN 61000-3-2, EN 61000-3-3 Applicable under the following conditions The maximum length of all cabling and wiring connected to the product must be less than 2.5 m.Shielded cables are being used when using the SIGNAL I/O.The high-voltage test lead TL31-TOS is in use.Electrical discharges are applied only to the EUT.
Safety *1 Complies with the requirements of the following directive and standards. Low Voltage Directive 2014/35/EU *2, EN 61010-1 (Class I *5 , Pollution Degree 2 *6)

*1 Does not apply to specially ordered or modified products.
*2 Only on models that have CE/UKCA marking on the panel.
*3 This is a Class A instrument. This product is intended for use in an industrial environment. This product may cause interference if used in residential areas. Such use must be avoided unless the user takes special measures to reduce electromagnetic emissions to prevent interference to the reception of radio and television broadcasts.
*4 This is a Group 1 instrument. This product does not generate and/or use intentionally radio-frequency energy, in the form of electromagnetic radiation, inductive and/or capacitive coupling, for the treatment of material or inspection/analysis purpose.
*5 This is a Class I instrument. Be sure to ground this product’s protective conductor terminal. The safety of this product is guaranteed only when the product is properly grounded.
*6 Pollution is addition of foreign matter (solid, liquid or gaseous) that may produce a reduction of dielectric strength or surface resistivity. Pollution Degree 2 assumes that only non-conductive pollution will occur except for an occasional temporary conductivity caused by condensation.

Get Adobe' Reader'

To view the pdf files, the Adobe Reader from Adobe Systems is required.

Researched by Shared Research Inc.
about Nihon Denkei Co., Ltd.