CONTENTS CONTENTS

Products

The products listed here are for overseas markets, not for Japan domestic.

ANRITSU METER Handheld Thermometer Model HR Series HR-1350K / HR-1350E

High function model. Can output measured values

Measurement Range
K -200 to 1370°C / E -200 to 800°C
 
HOLD
The thermometer stops measurement by the HOLD key and holds the present readings.
 
Auto power-off
The thermometer will turn the power off automatically after 5minutes.
 
Resolution change
Press the [RESO] key to change the resolution of readings.
 
P/V Hold
The peak (Max.) value and the valley (Min.) value can be displayed by the P/V HOLD key on the sub-display.
 
Backlight
The backlight will operate by the Backlight key and the LCD display will be visible even in a dark place.
 
Calibration
The displayed value can be adjusted the CAL SET key.
 
Alarm output
The thermometer gives an alarm when readings go beyond the high or low limits.

Made in Japan
Large easy-to-read LCD display

Handheld Thermometer Model HR Series

Model:

HR-1350K / HR-1350E

Maker:

ANRITSU METER CO., LTD.

Features

  • Accuracy : ±(0.05% of reading + 0.2℃ at over 0℃)
  • Resolution : 0.1℃ or 1℃ (Changeable)
  • Analog output : 1mV/℃ or 10mV/℃ (Changeable)
  • Alarm output function
  • Dimensions : 82(W)×166(H)×36(D)
  • Weight : 350g (including batteries)
  • Power supply : 4AA Alkaline dry batteries or AC adaptor
  • Battery life : approx.600hours
  • Accessories : Test report, Soft case, Hand strap, 4AA Alkaline batteries, Manual, Alarm/Analog output cable

Line up

Model

HR-1100
HR-1150
HR-1650

HR-1200
HR-1250
HR-1750
HR-1300
HR-1350
HR-1301
HR-1351
HR-1400
HR-1450
HR-1500
HR-1550
Display LCD  LED LCD 
Waterproof specifications/Equivalent to IPX5 ✔ ✔ – – ✔ –
Input connector ASP model:HR-1*0* /ANP model:HR-1*5*
Input type Thermocouple Input: Type E, K (Not switchable)
Channel 1ch
Signal source resistance MAX.1 Kohm
Measurement Range 1°C Resolution E -200~800°C
K -200~1370°C
0.1°C Resolution E -104.9~504.9°C(Switches automatically to 1°C resolution range when messured temperatures fall outside the 0.1°C resolution range.)
K -104.9~504.9°C(Switches automatically to 1°C resolution range when messured temperatures fall outside the 0.1°C resolution range.)
Measurement Accuracy 1°C Resolution 0°C~ ±(0.1% of reading +1°C)
~0°C ±(0.5% of reading +1°C)
0.1°C Resolution 0°C~ ±(0.05% of reading + 0.2°C)
~0°C ±(0.15% of reading + 0.2°C)
Reference junction compensation accuracy ±0.2°C at 25°C±10°C 

Temperature Coefficient*1

HR-1*1*:±0.02×Δt°C/HR-1*5*:±0.03×Δt°C (Add the value obtained by multiplying the excess temp.Δt[°C] by a coefficient to the total accuracy. e.g. ±0.3°C is added at room temp. of 0°C or 50°C)
Operation Environment 0 to 50°C, 0 to 80%RH (Non condensing)
Storage Environment -20 to 50°C, 0 to 85%RH (Non condensing)
Battery life approx.900h approx.900h approx.600h approx.20h*2 approx.300h approx.550h
Power supply(Dry battery) 4 AA Alkaline batteries
Power supply(AC power) – –

available

– available
Sampling rate approx.200ms
Linearlizer Digital Linearizer method (Compliant with IEC 60584-1 (2013))
Dimensions(mm) approx. 82(w)×166(H)×36(D)  [Protruded parts excluded]
Weight approx. 350g [Include Batteries]
Supplied accessories Manual,Test report,Soft case, Hand strap,4 AA Alkaline batteries
Optional accessories – – Alarm output cable – Communication cable,software
  Analog output cable – –
Alarm Output – –

✔ TABLE1

✔ TABLE1 –

‐

Analog Output – – – ✔ TABLE2

‐

‐

Memory function – – – – – ✔ TABLE3
Compatible standards(CE) EMC:EN61326-1:2013, EN 61326-2-1:2013 classA Table2( Industrial)

RoHS:IEC EN 63000:2018

*1 Only when the temp. exceeds 25°C ±10°C
*2 Approx. 400h when the analog output is switched always off.

Videos

Specifications

Model HR-1350
Display LCD
Waterproof specifications –
Input connector ANP
Input type Thermocouple Input: Type E, K
Channel 1ch
Signal source resistance MAX.1k ohm
Measurement Range 1°C Resolution E -200 to 800°C
K -200 to 1370°C
0.1°C Resolution E

-104.9 to 504.9°C

(Switches automatically to 1°C resolution range when messured temperatures fall outside the 0.1°C resolution range.)

K

-104.9 to 504.9°C

(Switches automatically to 1°C resolution range when messured temperatures fall outside the 0.1°C resolution range.)

Measurement Accuracy 1°C Resolution 0°C to ± (0.1% of reading +1°C)
to 0°C ± (0.5% of reading +1°C)
0.1°C Resolution 0°C to ± (0.05% of reading + 0.2°C)
to 0°C ± (0.15% of reading + 0.2°C)
Reference junction compensation accuracy ±0.2°C at 25°C±10°C
Temperature Coefficient*1

±0.03×Δt°C (Add the value obtained by multiplying the excess temp.Δt[°C] by a coefficient to the total accuracy. e.g. ±0.5°C is added at room temp. of 0°C or 50°C)

Operation Environment

0 to 50°C, 0 to 80%RH (Non condensing)

Storage Environment

-20 to 50°C, 0 to 85%RH (Non condensing)

Battery life approx.600h
Power supply(Dry battery) 4 AA Alkaline batteries
Power supply(AC power) option available
Sampling rate approx.200ms
Linearlizer Digital Linearizer method (Compliant with JIS C1602-2015)
Dimensions(mm) approx. 82(w)×166(H)×36(D)  [Protruded parts excluded]
Weight approx. 350g [Include Batteries]
Supplied accessories Manual,Test report,Soft case,
Hand strap,4 AA Alkaline batteries
Optional accessories Alarm output cable
Alarm output TABLE1
Compatible standards

EMC:EN61326-1:2013, EN 61326-2-1:2013 classA Table2( Industrial)

RoHS:IEC EN 63000:2018

*1 Only when the temp. exceeds 25°C±10°C