CONTENTS CONTENTS

Products

The products listed here are for overseas markets, not for Japan domestic.

ANRITSU METER Compact Thermologger AM-9000 Series AM-9200

Data recorded by simple operation can be transferred to PC with dedicated software for analysis processing.

Measurement Range
E -200~800℃ / K -200~1370℃ / J -200~1100℃ / T -200~400℃ / R 0~1760℃
 
Channel
6ch (ANP connector (both poles are coper))
 
Input Type
Thermocouple (multi-input : Type E,K,J,T,R)
Voltage input
 
Interface
USB2.0(microーB)
 
Interval
(The following numbers are time of interval that can be set.)
less than 1s : 0.2s ~ 0.99s Over 1s : 1s ~ 99m 59s

Compact Thermologger AM-9000 Series

Model:

AM-9200

Maker:

ANRITSU METER CO., LTD.

Features

  • Multi-input type. Type K, E, J, T, R changeable
  • Number of inputs : 6ch
  • Accuracy : ±0.5℃ (at 0.0~200.0℃)
  • Measurement interval : 0.2sec to 99min 59sec
  • Memory capacity : 120,000 data
  • Dimensions : 92(W)×155(H)×36(D), weight : 410g
  • Power supply : 4 AA Alkaline batteries or AC adaptor
  • Accessories : Test report, Soft case, Hand strap
  • 4AA Alkaline batteries, Manual, communications cable

Specifications

Model AM-9200
Number of inputs 6ch : ANP connector (Miniature Plug)
Input connector ANP connector (both poles are coper)
Input type Temperature:Multi-input (E・K・J・T・R)
DC Voltage:V
Measurement range Temperature: TABLE Ⅰ  DC Voltage : ±20.000V
Resolution Temperature: 0.1℃/1℃  DC Voltage : 0.001V
Measurement accuracy Temperature: TABLE Ⅱ  DC Voltage : ±(0.05% of reading +0.003V)
Memory capacity 120,000 data
Interface USB2.0 (micro-B)

Dimensions
*Protruded parts excluded

92(W)×155(H)×36(D)
Weight(Including dry batteries) approx.410g
Reference junction
compensation accuracy
±0.5℃(at 25℃±10℃)
Temperature coefficient

± (0.04 × Δt ℃) *1

Time interval

0.2 second to 99 minutes 59 seconds

Signal source resistance 500Ω or less (measurement interval 1 second or more)/100Ω or less (measurement interval less than 1 second)
Safe excitation voltage Between input pins on same channels :±30VDC,Between input pins on adjacent channels : 150Vp-p
Linearlizer Digital Linealizer method (Compliant with JIS C1602-2015)
Display Graphic liquid crystal display, 128×64 dots
Environment operational
Environment storage

Operation : 0〜40℃、15〜80%RH (Non condensing)  
Storage : -10〜50℃、10〜85%RH (Non condensing)

Battery life TABLE Ⅲ
Power supply 4 AA Alkaline batteries(Supplied accessories) or AC-Adaptor(Optional accessories)
Supplied accessories Manual, Test report, Softcase, Communications cable, 4 AA Alkaline batteries
Optional accessories Special data analysis software, AC-Adaptor, Carrying case
Compatible standards CE Marking, RoHS2

*1 When the ambient temperature is other than from 15℃ to 35℃, “the excess temperature from the ambient temperature Δt℃” x 0.04 is added to the measurement accuracy.

TABLE Ⅰ : Measurement Range

Thermocouple type Measurement range
E -200 to 800°C
K -200 to 1370°C
J -200 to 1100°C
T -200 to 400°C
R 0 to 1760°C

TABLE Ⅱ : Measurement accuracy

In the environment of 25 ℃ ± 10 ℃, when the measurement interval is 1 second or more without including the reference junction compensation accuracy.

Thermocouple type Accuracy compensation range Measurement accuracy
E,K,J,T -200.0~-100.1℃ ±1.5℃
-100.0~-0.1℃ ±1.0℃
0.0~200.0℃ ±0.5℃
200.1~500.0℃ ±0.7℃
500.1℃ or more ±1.0℃
R 0.0℃~500.0℃ ±2.0℃
500.1℃ or more ±2.5℃

*If the measurement interval is less than 1 sec, add ± 0.5 ℃ at type E ・ K ・ J ・ T and add ± 1.0℃ at type R

TABLEⅢ 12ch model standard time for continuous use (at 25 ℃, using alkaline AA batteries)

Setting interval 0.2sec 1sec 10sec
AM-9000 Series approx. 50 hours approx. 55 hours approx. 140 hours

* If the number of used channels is less than 12, the usable time will be longer.