This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.
Products
The products listed here are for overseas markets, not for Japan domestic.
Test Frequency 10Hz-50MHz
The MICROTEST 6632 Impedance Analyzer offers a wide frequency range of 10Hz to 50MHz for measuring the impedance characteristics of components and materials. The measured parameters include |Z| (impedance), |Y| (admittance), θ (phase angle), X (reactance), R (series/parallel resistance), G (conductance), B (Susceptance), L (inductance), D (dissipation factor), Q (quality factor), DCR (DC resistance), C (capacitance), ESR (equivalent series resistance), ε (relative permittivity), and μr (relative permeability). It supports both Meter mode and Sweep Analysis mode, allowing measurement results to be displayed in numeric or graphical formats. It supports LAN/USB Host/USB Device/RS-232/GPIB (Optional) communication interfaces, making it compatible with automation equipment or human-machine interfaces (PLC).
Impedance Analyzer 6632 Series
Model:
6632-1, 6632-3, 6632-5, 6632-10, 6632-20, 6632-30, 6632-50
Maker:
MICROTEST
Features
- Frequency range: DC, 10Hz to 1/3/5/10/20/30MHz/50MHz
- Basic accuracy up to ±0.08% (typical ±0.05% )
- Automatic Level Control (ALC)
- Output impedance 25Ω/100Ω, switchable
- Support meter mode and list mode, sweep mode
- Built-in DC Bias voltage ±12V
- Ultra-high measuring speed < 3ms
- Open circuit/short circuit/load correction function
- Comparator function
- BIN classification function
- Built-in the Permeability-μr
- Built-in the Dielectric constant-εr
Line up
| 6632 Series | |||||||
|---|---|---|---|---|---|---|---|
| Model | 6632-1 | 6632-3 | 6632-5 | 6632-10 | 6632-20 | 6632-30 | 6632-50 |
| Frequency | 10Hz-1MHz | 10Hz-3MHz | 10Hz-5MHz | 10Hz-10MHz | 10Hz-20MHz | 10Hz-30MHz | 10Hz-50MHz |
To measure high-frequency components, use a 6632 impedance analyzer.
Specifications
| Model | 6632 | |
| Frequency Range | 10Hz-50MHz | |
| Frequency Setting | Continuity | |
| Frequency Resolution | 100mHz, 6-bit Frequency Input | |
| Frequency Output Accuracy | ±0.01% | |
| Basic Accuracy | ±0.08% (typical ±0.05%) | |
| AC Drive Level | Test Signal Voltage Level |
10mV – 2Vrms(100Ω) 10mV – 1Vrms(25Ω) |
| Voltage Minimum Resolution | 1mV | |
| Accuracy |
ALC ON:6% * Voltage ±2mV ALC OFF:10% * Voltage ±2mV |
|
| Test Signal Current Level | 200µA – 20mArms | |
| Current Minimum Resolution | 10µA | |
| Accuracy |
ALC ON:6% * Current ± 20µA ALC OFF:10% * Current ± 20µA |
|
| DC Drive Level | 1V (fixed) | |
| Output Impedance | 25Ω/100Ω (switchable) | |
| DC Bias Voltage |
Built-in ±12V (100Hz-50MHz) |
External Voltage Bias (±12V/±40V/±600V/±2000V) |
| DC Bias Current |
Option DC Bias Current Source (20A/60A/120A/320A) |
External Voltage Bias (100mA) |
| Measurement Mode | Meter mode, list mode, sweep mode, and optional equivalent circuit analysis function (S model) | |
| Measurement Parameters |
|Z|、|Y|、θ、X、R、G、B、L、D、Q、DCR、C、 ESR、εr 、μr |
|
| Test Time (Fastest) | < 3mS | |
Measurement Parameters
| Parameters | Range |
| |Z| | 0.000 mΩ to 9999.99 MΩ |
| R, X | ± 0.000 mΩ to 9999.99 MΩ |
| |Y| | 0.00000 μS to 999.999 kS |
| G, B | ± 0.00000 μS to 999.999 kS |
| θRAD | ± 0.00000 to 3.14159 |
| θDEG | ± 0.000° to 180.000° |
| Cs, Cp | ± 0.00000 pF to 9999.99 F |
| Ls, Lp | ± 0.00 nH to 9999.99 kH |
| D | 0.00000 to 9999.99 |
| Q | 0.00 to 9999.99 |
| Δ | ± 0.00% to 9999.99% |
| Rdc | 0.00 mΩ to 99.9999 MΩ |
| εr’ εr” | 0 to 100000 |
| μr’ μr” | 0 to 100000 |
System
| Measurement Circuit | Series/Parallel |
| Correction | Open Circuit/Short Circuit/Load correction |
| Cable Compensation | 0 / 0.5 / 1 /2 (m) |
| BIN | ABS、ΔABS、Δ%、OFF |
| Comparator | 9 |
| List Mode | 50 groups of Multi-steps setting (Each group contains up to 15 steps) |
| Bulit-in Storge | 100 sets LCR setting documents, 50 groups of list mode setting |
| USB Host Storge | LCR setting documents, list mode setting document, BMP graphics, Sweep screen and test result data |
| Trigger Test | Auto, manual, RS-232, GPIB, Handler |
| Power Supply | Voltage:90-264Vac |
| Frequency:47-63Hz | |
| Power consumption | 30VA |
| Environment | Temperature:10℃-40℃ |
| Humidity:20-90%RH | |
| Dimension | 336x147x340 mm (W*H*D) |
| Weight | 3.95 Kg |
| Interface | RS-232、Handler、LAN、USB Host、USB Device |
| Display | Color Screen, 7″ TFT (800*480) |