This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.


ITEMLIST
Products Infomation

Line Impedance Stabilization Network TNW-403D2
TNW-403D2 is a Line impedance stabilization network(LISN) used to measure interference voltage leaking from the interference sources due to the power supply line in the frequency range of 9kHz to 30MHz, in compliance with standards such as CISPR (International Special Committee on Radio Interference).
TNW-403D2 is designed for single-phase and can measure Δ and V-connection.
Additionally, by registering the correction factors of this device into radiation receiver manufactured by Kyoritsu Electronic Industry (KNM-2401, KNM-2402, KNM-2403, and KNM-S243), automatic addition can be performed through key operations, enabling direct reading measurement of interference voltage.
Line Impedance Stabilization Network TNW-403D2
Model:
TNW-403D2
Maker:
Kyoritsu Technology Corp.
Specification
Frequency range | 150kHz to 30MHz |
Applicable standards | CISPR 16-1-2 |
Test standards | CISPR 11/ 13/ 14/ 15/ 22, PSE |
Number of phase | Single |
Circuit type | 150Ω (⊿ and V connection) |
RF connector | 50Ω(BNC female) |
Power Line Rating | AC/DC 125V, 15A |
Maximum Rating | AC/DC 250V, 15A |
Standard accessories | Coaxial cable (BNCP-3D2W-BNCP, 2m) ・・・・・1pc.
AC power cable (3m) ・・・・・・・・・・・・ 1pc. |
Size / Weight | 300(W)x130(H)x160(D)mm, approx.4kg |
Available EMI test receiver | TNM-2401, TNM-2402 |
Built-to-Order Model | |
Options | High Pass Filter TFL-009, TFL-009D |
Precaution for use
1.Using a leakage current breaker on the power line may cause the breaker to trip.
Leakage current of AC100V, 50/60Hz, 400mA flows through the line.
[Countermeasures]
1) Use an isolation transformer.
2) Use a power line without a breaker installed.
2.When switching the measurement circuit, pulses are generated. When combined with a spectrum analyzer, there is a risk of damaging the input circuit of the spectrum analyzer.
[Countermeasures]
1) Using a high-pass filter TFL-009 or TFL-009D.
2) Switching after disconnecting the coaxial cable.
3) Following the requirements of the CISPR standard, inserting a 10dB attenuator inside or outside the lime impedance stabilization network to match the measuring terminals of the instrument to 50Ω. This model supports external attachment for this purpose.