CONTENTS CONTENTS

Products

The products listed here are for overseas markets, not for Japan domestic.

ESPEC Bench-Top Type Temperature (& Humidity) Chambers SH / SU Series

Compact but realizing efficient environmental for testing with high-performance networking

The bench-top-sized thermo-hygrostat and small-sized environmental testing equipment are compact but have full-scale environmental testing performance.
Temperature change rate of 5.0°C per minute is added to the temperature (humidity) range combined with internal capacity of 22.5L and 64L. A total of 12 models are available.
The newly developed N instrumentation incorporates a color touch panel, offering operability and visibility similar to those of the well-reputed Platinus J series.

ESPEC Bench-Top Type Temperature (& Humidity) Chamber

Model:

SH-222,

Maker:

ESPEC CORP

Features

  • Compact enough for desktops
  • Stackable for enhanced space efficiency
  • Instrumentation Interlock I/O Terminals
  • Three-way access

Line up

Model Capacity Temperature range Humidity range Temperature change rate (pull down)
Temperature and humidity chamber SH-222 22.5 L -20 to +150°C 30%rh to 95%rh
(See the “Temperature and humidity control range chart” on P.10)
2.1°C/min
SH-242 -40 to +150°C
SH-262 -60 to +150°C
SH-642 64 L -40 to +150°C 1.7°C/min
SH-662 -60 to +150°C
SH-242-5 22.5 L -40 to +150°C 5°C/min
Temperature chamber SU-222 22.5 L -20 to +150°C – 2.1°C/min
SU-242 -40 to +150°C
SU-262 -60 to +150°C
SU-642 64 L -40 to +150°C 1.7°C/min
SU-662 -60 to +150°C
SU-242-5 22.5 L -40 to +150°C 5°C/min

Specifications

SH Series Specifications

Model SH-222 SH-242 SH-262 SH-642 SH-662 SH-242-5
Temp. performance System Balanced Temperature & Humidity Control system (BTHC system)
Temp. range -20 to +150°C -40 to +150°C -60 to +150°C -40 to +150°C -60 to +150°C -40 to +150°C
Temp. fluctuation ±0.3°C (-20/-40/-60 to +100°C) / ±0.5°C (+100.1 to +150°C)
Temp. variation in space 2.5°C (-20/-40/-60 to +100°C) / 4.0°C (+100.1 to +150°C)
Heat up rate 3.2°C/min 2.9°C/min 5.0°C/min
Pull down rate 2.1°C/min 1.7°C/min 5.0°C/min
Heat up time From -20 to +150°C
within 55 min.
From -40 to +150°C
within 60 min.
From -60 to +150°C
within 70 min.
From -40 to +150°C
within 70 min.
From -60 to +150°C
within 80 min.
From -40 to +150°C
within 40 min.
Pull down time From +20 to -20°C
within 20 min.
From +20 to -40°C
within 50 min.
From +20 to -60°C
within 70 min.
From +20 to -40°C
within 60 min.
From +20 to -60°C
within 90 min.
From +20 to -40°C
within 20 min.
Lowest attainable temp. -20°C -40°C -60°C -40°C -60°C -40°C
Humid. performance Humid. range 30 to 95%rh
Humid. fluctuation ±3.0%rh
Construction Heater Nichrome strip wire heater
Humidifier Stainless steel cartridge heater
Refrigeration unit System Mechanical single-stage Mechanical cascade refrigeration system Mechanical single-stage
Cooler Plate fin cooler
Refrigerator capacity 400W [Unit 1: 400W x1, Unit 2: 400W x1] 400W
Refrigerant R-404A R-404A, R-508A (R-23 for 100V) R-404A, R-23
Capacity 22.5 L 64 L 22.5 L
Chamber total load resistance 20 kg
Inside dimensions mm W300 × H300 × D250 W400 × H400 × D400 W300 × H300 × D250
Outside dimensions mm W440 × H690 × D696 W440 × H690 × D786 W540 × H730 × D921 W440 × H690 × D786
Weight 83 kg (78 for 100V) 105 kg 130 kg 106 kg
Utility requirements Allowable ambient conditions +5 to +35°C
Power supply 100V AC 1φ 50/60Hz 11.3 A 15.0 A 17.5 A 16.7 A
115V AC 1φ 60Hz (NEC) 12.8 A – – –
200V AC 1φ 50/60Hz – 14.0 A 14.5 A 10.6 A
220V AC 1φ 50/60Hz 5.4 A 13.5 A 14.0 A 9.3 A
230V AC 1φ 50Hz 5.3 A 13.5 A 14.0 A 9.2 A
Noise level Between 42 and 50 dB Between 42 and 54 dB Between 48 and 53 dB Between 42 and 54 dB
Exhaust heat quantity 3500 kJ/h 4200 kJ/h 5040 kJ/h 5700 kJ/h

SU Series Specifications

Model SU-222 SU-242 SU-262 SU-642 SU-662 SU-242-5
Temp. performance System Balanced Temperature Control system (BTC system)
Temp. range -20 to +150°C -40 to +150°C -60 to +150°C -40 to +150°C -60 to +150°C -40 to +150°C
Temp. fluctuation ±0.3°C (-20/-40/-60 to +100°C) / ±0.5°C (+100.1 to +150°C)
Temp. variation in space 2.5°C (-20/-40/-60 to +100°C) / 4.0°C (+100.1 to +150°C)
Heat up rate 3.2°C/min 2.9°C/min 5.0°C/min
Pull down rate 2.1°C/min 1.7°C/min 5.0°C/min
Heat up time From -20 to +150°C
within 55 min.
From -40 to +150°C
within 60 min.
From -60 to +150°C
within 70 min.
From -40 to +150°C
within 70 min.
From -60 to +150°C
within 80 min.
From -40 to +150°C
within 40 min.
Temp. extreme achievement time Pull down time From +20 to -20°C
within 20 min.
From +20 to -40°C
within 50 min.
From +20 to -60°C
within 70 min.
From +20 to -40°C
within 60 min.
From +20 to -60°C
within 90 min.
From +20 to -40°C
within 20 min.
Lowest attainable temp. -20°C -40°C -60°C -40°C -60°C -40°C
Heater Nichrome strip wire heater
Construction Refrigeration unit System Mechanical single-stage refrigeration system Mechanical cascade refrigeration system Mechanical single-stage refrigeration system
Cooler Plate fin cooler
Refrigerator capacity 400W [Unit 1: 400W x1, unit 2: 400W x1] 400W
Refrigerant R-404A R-404A, R-508A (R-23 for 100V type) R-404A, R-23
Capacity 22.5 L 64 L 22.5 L
Chamber total load resistance 20 kg
Inside dimensions mm W300 × H300 × D250 W400 × H400 × D400 W300 × H300 × D250
Outside dimensions mm W440 × H625 × D696 W540 × H665 × D921 W440 × H625 × D786
Weight 78 kg (73 for 100V type) 100 kg 123 kg 101 kg
Utility requirements Allowable ambient conditions +5 to +35°C
Power supply 100V AC 1φ 50/60Hz 9.3 A 15.0 A 17.5 A – 16.7 A
115V AC 1φ 60Hz (NEC) 11.0 A – – – –
200V AC 1φ 50/60Hz – 14.0 A 14.5 A 10.6 A
220V AC 1φ 50/60Hz 4.5 A 13.5 A 14.0 A 9.3 A 9.3 A
230V AC 1φ 50Hz 4.4 A 13.5 A 14.0 A 9.2 A 9.2 A
Noise level Between 42 and 50 dB Between 42 and 54 dB Between 48 and 53 dB Between 42 and 54 dB
Exhaust heat quantity 3500 kJ/h 4200 kJ/h 5040 kJ/h 5700 kJ/h

Get Adobe' Reader'

To view the pdf files, the Adobe Reader from Adobe Systems is required.

Researched by Shared Research Inc.
about Nihon Denkei Co., Ltd.