CONTENTS CONTENTS

Products

The products listed here are for overseas markets, not for Japan domestic.

ANRITSU METER High Superior for Extremely Small Surface Probes S-4 Series S-423K-01-1-TPC1-ANP

Highest levels of accuracy,response time and durability.

Model S series probes are high-performance temperature sensors developed using Anritsu Meter's full range of proprietary technologies to achieve the highest levels of accuracy, response time, and durability.
The head of the SX is tilted士4°, enabling stable measurement at the time of contact.

High Superior for extremely small Surface Probes S-4 Series

Model:

S-423K-01-1-TPC1-ANP

Maker:

ANRITSU METER CO., LTD.

Features

  • Small head:φ6×7mm
  • Temperature range :-50 to 800°C
  • Tolerance :±2.5°C (at 100°C surface)
  • Response time :1s
  • Durability :More than 60,000 contacts (at 300°C surface)

Specifications

Model number*1 S-41* S-42*
Thermocouple type Type E or K
Temp. range*2 -50 to 300°C -50 to 800°C
Tolerance*3 -50℃ ±2.5°C ±2.5°C
0°C ±2.5°C ±2.5°C
100°C ±2.5°C ±2.5°C
200°C ±2.5°C ±2.5°C
300°C ±2.5°C ±2.5°C
400°C – ±3.0°C
500°C – ±3.8°C
600°C – ±9.0°C
700°C – ±10.5°C
800°C – ±16.0°C
Tolerance calculation method
t:Temperature(°C)
At within temperature range
1)-50°C ≦ t < -40°C : ±2.5°C 
2)0°C ≦ t ≦ 500°C : ±2.5°C or ± (0.0075 × | t |)°C, whichever is greater
3)500°C < t ≦ 700°C : ±(0.015×|t|)°C
4)700°C < t ≦ 800°C : ±(0.02×|t|)°C
Calibration points of test report 100, 200, 300°C 100, 300, 500°C
Response time*4 0.1 to 0.15 s(63.2%) 0.3 to 0.4 s(90%) 1 s(99%)
Durability*5 More than 100,000 contacts at 300°C More than 60,000 contacts at 800°C
allowable tilt ±5°
Head material Liquid Crystal Polymer Ceramic
Pipe material Stainless(SUS304)
Contact strip material Heat resistant alloy t=50μm
Grip material Polyacetal
Repair*6 Repairable

*1 The asterisk (*) is replaced by the number of the model name you selected. The model number after thermocouple type is omitted.
*2 The operating temperature limit is determined by the allowable temperature limit of the sensor head contacts the measurement target. Note that the operatingtemperature limit is not the same as the allowable temperature limits of the grip, cable, and plug.
*3 Tolerance is available at 0°C or above within the operating temperature on a stationary flat and smooth metal surface. For tolerance at temperature not listed in the above table, please refer to calculation method of tolerance.
*4 The response time is the time required to detect the true value on a flat and smooth metal surface.
*5 Number of contacts enabling measurement within the tolerance range on a flat and smooth metal surface.
*6 Please refer to page repair.

Get Adobe' Reader'

To view the pdf files, the Adobe Reader from Adobe Systems is required.

Researched by Shared Research Inc.
about Nihon Denkei Co., Ltd.