This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.
Products
The products listed here are for overseas markets, not for Japan domestic.
Multi-purpose temperature probes featuring a flexible tape shape.
Model STN series probes are multi-purpose temperature sensors featuring a flexible tape shape. They can be glued to the surfaces of measurement targets for use in a range of applications. Specific components may be combined to customize products for the needs of a customer's specific application.
Flat Leaf Probes (tape replaceable type) Model STN Series
Model:
STN-34K-015
Maker:
ANRITSU METER CO., LTD.
Features
- Cable length : 1m (Sheath material : Silicone rubber)
- Terminal 1 : Plug for STN tape part
- Terminal 2 : Miniature plug
Specifications
| Model number*1 | STN-*1 | STN-*2 | STN-*3 | STN-*4 | ||
| Thermocouple type | Type E or K | |||||
| Temp. range*2 | -50 to 210°C | -50 to 210°C | Tape length ≦ 150mm : -50 to 400°C Tape length > 150mm : -50 to 350°C |
|||
| Tolerance*3 | -50°C ≦ t < -40°C | ±3.5°C | ±3.5°C | ±3.5°C | ±3.5°C | |
| 0°C | ±2.5°C | ±2.5°C | ±2.5°C | ±2.5°C | ||
| 100°C | ±2.5°C | ±2.5°C | ±2.5°C | ±2.5°C | ||
| 200°C | ±2.5°C | ±2.5°C | ±2.5°C | ±2.5°C | ||
| 300°C | – | – | ±2.5°C | ±2.5°C | ||
| 400°C | – | – | ±3.0°C | ±3.0°C | ||
| Tolerance calculation method t:Temperature(°C) |
At within temperature range -40°C ≦ t ≦ 400°C:±2.5°C or ± (0.0075 × | t |)°C, whichever is greater |
|||||
| Response time*4 | 2s | 8s | 2s | 8s | ||
| Durability*5 | S(Can not evaluate because of special-application) | |||||
| Repair*6 | Non-repairable | |||||
*1 The asterisk (*) is replaced by the number of the model name you selected. The model number after thermocouple type is omitted.
*2 The operating temperature limit is determined by the allowable temperature limit of the sensor head contacts the measurement target. Note that the operatingtemperature limit is not the same as the allowable temperature limits of the grip, cable, and plug.
*3 Tolerance is defined in the operating temperature range of at the internal temperature with sufficient heat capacity. For tolerance at temperature not listed in the above table, please refer to calculation method of tolerance.
*4 The response time given for our temperature measuring probes is the time required to detect 99% of the true value when measuring the temperature of boiling water or vapor.
*5 Can not evaluate because of special-application.
*6 Please refer to page repair.