KIKUSUI PLZ-5W Series

KIKUSUI - Multifunctional Electronic Load PLZ-5W Series -

Price : [Contact Us]
 

The Kikusui PLZ-5W Series electronic loads feature a 4 μs rise time and a maximum slew rate of 60 A/μs, ensuring rapid transient response. They support multiple operation modes—CC, CR, CV, CP, and ARB—and include a high-resolution color LCD and numeric keypad for enhanced usability.

Multifunctional Electronic Load

Model:

PLZ-5W Series

Maker:

KIKUSUI

Download Materials

Line-up

Image SKU Rated operating voltage [V] Rating current [A] Rating Power [W]
PLZ205W 0.25 to 150 40 200
PLZ405W 0.25 to 150 80 400
PLZ1205W 0.25 to 150 240 1200

General Specification

Model PLZ205W PLZ405W PLZ1205W
Input voltage range/ Input frequency range 100 Vac to 240 Vac (90 Vac to 250 Vac) single phase, continuous / 47 Hz to 63 Hz
Power consumption 50VAmax 50VAmax 85VAmax
Inrush current (peak value) 45A Peak
Operating temperature range / Operating humidity range 0 °C to 40 °C (32 °F to 104°F) / 20%rh to 85%rh (no condensation)
Storage temperature range / Storage humidity range -20 °C to 70 °C (-4 °F to 158°F) / 90%rh or less (no condensation)
Installation location Indoor use, altitude of up to 2000 m, overvoltage category II.
Between primary and input terminals 500Vdc, 30 MΩ or more (70%rh or less)
Between primary and chassis
Between input terminals and chassis
Between primary and input terminals No abnormalities at 1500 Vac for 1 minute.
Between primary and chassis No abnormalities at 1500 Vac for 1 minute.
Between input terminals and chassis No abnormalities at 750 Vac for 1 minute.
Dimensions Unit: mm (inches) 214.5 (8.45)W×124 (4.88)H×400 (15.75)Dmm(inches) 429.5 (16.91)W×128 (5.04)H×400 (15.75)Dmm(inches)
Weight Approx. 7 kg (15.4 lb.) Approx. 7.5 kg (16.5 lb.) Approx. 14 kg (30.9 lbs)
Accessories Power cord, Rear-panel load input terminal cover, Load input terminal screw set (2 sets), Screws for the rear-panel load input terminal cover (2 pcs.), Front-panel load input terminal cover, Front-panel load input knob set, External control connector kit, Setup Guide, CD-ROM, Quick Reference, Safety Information
Electromagnetic compatibility (EMC) *1 *2 Complies with the requirements of the following directive and standards. EMC Directive 2014/30/EU, EN 61326-1 (Class A*3), EN 55011 (Class A*3, Group 1*4), EN 61000-3-2, EN 61000-3-3 Applicable under the following conditions.The maximum length of all cabling and wiring connected to the PLZ-5W must be less than 3 m.
Safety *1 Complies with the requirements of the following directive and standards. Low Voltage Directive 2014/35/EU *2 EN 61010-1 (Class I *5, Pollution Degree 2 *6)

*1 Does not apply to specially ordered or modified PLZ-5Ws. *2 Limited to products that have the CE/UKCA mark on their panels. *3 This is a Class A equipment. This product is intended for use in an industrial environment. This product may cause interference if used in residential areas. Such use must be avoided unless the user takes special measures to reduce electromag-netic emissions to prevent interference to the reception of radio and television broadcasts. *4 This is a Group 1 equipment. This product does not generate and/or use intentionally radio-frequency energy, in the form of electromagnetic radiation, inductive and/or capacitive coupling, for the treatment of material or inspection/analysis purpose. *5 This is a Class I equipment. Be sure to ground this product’s protective conductor terminal. The safety of this product is only guaranteed when the product is properly grounded. *6 Pollution is addition of foreign matter (solid, liquid or gaseous) that may produce a reduction of dielectric strength or surface resistivity. Pollution Degree 2 assumes that only non-conductive pollution will occur except for an occasional temporary con-ductivity caused by condensation.