This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.
ITEMLIST
Products Infomation
High-performance DC voltage current source/monitor capable of pulse source/measurement with a minimum pulse width of 50 μs
The 6241A/6242 is a DC voltage and current source/ monitor capable of 4½-digit generation and 5½-digit measurement with high accuracy of ±0.02 %.
The instrument has two slope linear sweep, linear sweep, fixed sweep and random sweep functions, and is also capable of pulse measurement with a minimum pulse width of 50 μs and low voltage/current measurement with a resolution of 1μ V/100 pA. Thus, it can be used for a wide range of applications for evaluation in R&D of semiconductors and other electronic components, or for characteristic tests in a production line.
The sink-enabled bipolar output, the individual HI/LO limiter function and the suspend function are effective in evaluation of batteries and power ICs.
DC Voltage Current Source/Monitor
Model:
6241A / 6242
Maker:
ADC CORPORATION
Features
Source and measurement function
Voltage/current source and voltage/current/resistance measurement can be selected by specifying the source and measurement functions.
Voltage/current source mode
There are four source modes; DC, pulse, DC sweep, pulse sweep. Then, the sweep modes are classified into four sweep types: fixed sweep, linear sweep, random sweep (user programmable sweep), two slope linear sweep (linear sweep with step value switching). The minimum pulse width is 50 μs. The minimum cycle is 2 ms, or 500 μs without measurement.
Individual settings of HI/LO limiters
In voltage/current source, the HI/LO limiter settings are very important. The 6241A/6242 has a function that can set the HI and LO limiters individually. Furthermore, for the voltage-limiter, both HI and LO limiters can be set homo-polar. This prevents a capacitor or a battery from being over-discharged. Also, it is suitable for evaluating devices such as LDs that are used at a constant current and do not tolerate reverse voltage application.
Suspend function
The 6241A/6242 can select from three output OFF statuses; STBY (output relay OFF), HiZ (output relay ON and high resistance), and LoZ (output relay ON and low resistance). Using this function can omit unnecessary relay ON/OFF action, which reduces deterioration of the throughput due to relay operation time and improves the life span of the relay. In addition, the setting of a suspend voltage (voltage in HiZ and LoZ status) prevents transient current sink that occurs at a connection of a voltage output device such as a battery.
Sample hold measurement
Sample hold measurement is available in the pulse mode and the pulse sweep mode. Measurement holds an input signal immediately before pulse falling edge, and A/D conversion is implemented for integration time of 100 μs. In this case, the setting measurement delay time is ignored.
Specification
6241A | 6242 | ||
Number of digits for generation | 4½ | ||
Output method | Bipolar | ||
Maximum output (top) | Voltage | ±32 V/0.5 A | ±6 V/5 A |
10 μV | |||
Minimum resolution (bottom) | Current | ±0.5 A/32 V | ±5 A/ 6V |
1 nA | |||
Number of digits for measurement | 5½ | ||
Accuracy (typical range) | 0.02 % | ||
Minimum measurement resolution | Voltage | 1 μV | |
Current | 100 pA | ||
Maximum measurement resistance / Minimum resolution | 1.6 GΩ/2 μΩ | 304 MΩ/0.2 μΩ | |
Pulse application / Measurement | Available | ||
Minimum pulse width | 50 μs | ||
Interface | USB, GPIB |