This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.

ITEMLIST
Products Infomation

Enable everyone to easily evaluate reliability of PCA
The PCAS-1000B can be used even by a beginner to measure strain, and PCB strain speed and other measurements can be taken easily in accordance with IPC/JEDEC-9704A.* *IPC/JEDEC-9704A defines a method of evaluating strain when assembling, testing, and using a PCB. Excessive strain on a PCB can cause failure, so this is an effective means of evaluating quality and analyzing failure.
* The method of evaluating PCBs with strain gages is provided by IPC/JEDEC-9704A.
The method can be effectively for various purposes including quality evaluation and trouble analysis.
PCA Stress Measuring Set
Model:
PCAS-1000B
Maker:
Kyowa Electronic Instruments
Features
– Simply follow directions to seamlessly configure instruments, record data, and prepare reports.
– Automatically create reports optimal for evaluating PCBs.
– A single set is capable of 4-channel measurement, with a maximum of up to 16 channels.
Application
Video
Specifications
PCD-400B Sensor Interface
Measuring Targets | Strain gage and strain-gage transducers |
---|---|
Channels | 4 |
Input Form | Balanced differential input |
Synchronous Measurement | Up to 4 units by 16 channels |
Compatible Gage Resistance | Quarter bridge 2-wire system, 3-wire system: 120 Ω Half bridge system, full bridge system: 120 to 1000 Ω |
Input Connector | D-sub 37-pin connector |
Bridge Excitation | AC 2 Vrms |
Gage Factor | 2.00 fixed |
Balance Adjustment | Resistance: Within ±2% (±10 k ×10-6 strain) Capacitance: Within 5000 pF |
Balance Adjustment Methods | Resistance: Auto balance Capacitance: CST method (Capacitance self-tracking) |
Nonlinearity | Within ±0.1% FS |
Range | 200, 500, 1 k, 2 k, 5 k, 10 k, and 20 k ×10-6 strain – 7 steps Accuracy: within ±0.5% FS |
Frequency Response | DC to 200 Hz, deviation: Within ±10% |
Sampling Frequencies | Max. 10 kHz (Synchronous 4-unit sampling for 16 channels at 10 kHz) |
LPF | Transfer characteristic: 2nd order Butterworth Cutoff frequencies: 10, 30, 100 Hz, and FLAT (4 steps) Amplitude ratio at cutoff point: -3 ±1 dB Attenuation: (-12 ±1) dB/oct. |
AD Converter | 24 bits |
Holding Setting Condition | The range and balance adjustment value etc. are written to nonvolatile memory. |
TEDS | Reads information from TEDS-installed sensors. (Input adapters: UI-10A and UI-11A only) Channel name writing (If the manufacturer’s ID is Kyowa) |
Interfaces | USB2.0 (Conforms to High-speed USB standards.USB3.0 supported) |
Stability | Temperature Zero point: Within ±0.2 ×10-6 strain per ºC Sensitivity: Within ±0.05%/ºCTime Zero point: Within ±0.5 ×10-6 strain per 8h Sensitivity: Within ±0.15%/8h |
Withstand Voltage | 250 VAC for 1 minute between input and case 250 VAC for 1 minute between input channels (If the manufacturer’s ID is Kyowa) |
Operating Temperature | 0 to 40ºC |
Operating Humidity | 20 to 85% (Non-condensing) |
Vibration Resistance | ±29.42 m/s² (3 G) 5 to 200 Hz (12 cycles for each axis, 10 minutes/cycle) *unstacked |
Power Supply | 11 to 16 VDC Connector type: RM12BRD-4PH (Hirose) |
Current Consumption | 0.7 A or less (12 VDC) 0.6 A or less (15 VDC) |
Dimensions | 210 W x 35 H x 157.5 D mm (Excluding protrusions) |
Weight | Approx. 650 g |
Degree of Protection | IP30 (JIS C 0920 / IEC 60529) *Except when the one-touch type input adapter UI-55A is connected and when the adapter is not connected. |
*The specifications apply to the state in which the temperature has stabilized after a preheating time of 1 hour. |
UI-55A One-touch Type Input Adapter
Channels | 4 |
---|---|
Sensor Connection | One-touch type terminals |
Main Measuring Instruments | PCD-400B/430B EDX-11A/14A/15A (For full bridge only) |
Measuring Targets | 2-wire quarter bridge 3-wire quarter bridge Half bridge Full bridge Voltage Strain-gage transducer |
Dimensions | 105 W × 18.5 H × 35 D mm |
Weight | Approx. 90 g |