ECG-Kokusai DWX-15/DWX-15PD

Impulse Tester/Impulse Winding Tester DWX Series DWX-15/DWX-15PD

[Contact Us]
 
ECG KOKUSAI Co., Ltd. specializes in cutting-edge measurement hardware and inspection solutions, highlighted by their DWX and DWIT impulse winding testers. Notably, the DWX-05PD model includes Partial Discharge detection functions, crucial for assessing insulation integrity. Their innovative offerings with the DWX Series Impulse Testers — and the long-awaited 15 kV model is now available! Their product lineup, adhering to IEC60034 standards, also encompasses rotational direction testers and motor vibrophonic monitors.

Impulse Tester/Impulse Winding Tester

Model:

DWX-15/DWX-15PD

Maker:

ECG KOKUSAI

Specifications

Model DWX-15 DWX-15PD
Applied voltage/applied steps: *With 1 K Ω load 1000V~15000V
(200V steps)
Test impedance range 50μH or higher
Sampling speed / memory 8bit/10nano-sec.(100MHz) / 8192-Byte
Sampling ranges (WIDTH) -4、-3、-2、-1、0、1、2、3、4、5
10 ranges (0 is compatible with the DW Series)
Test input circuit
(Impedance)
Resistance division(5MΩ)
Screen display,
waveform display range
640 x 460 (VGA), 8.4-inch TFT color liquid crystal, 4-color display
512×256
Judgment method Comparison of master waveform with test waveform Judgment based on comparison of waveform area, waveform differential area and discharge quantity Winding insulation dielectric breakdown voltage test function
Master waveform memory capacity Body 196 types (14 types/14 pages)
Expansion 700 types with Compact Flash Memory (14 types/50 pages)
Microwave sensor antenna None Partial discharge electromagnetic sensing by narrow bandwidth flat batch antenna
Detection range None It is recommended that a distance of 30 cm be allowed from the test piece to the antenna when the device is installed.
Antenna dimensions/Length of cable between antenna and body None W = 80 mm, H = 80 MM, D = 10 mm,
cable length = 2 m
External interface Parallel I/O (Start, reset, OK, NG, Busy, mater switching, etc.)
RS-232C (Test control, test data, etc.)
Printer board (Hard screen copies, etc.)
Ethernet adapter (optional)
Accessories 15-M test cable: 1, power supply cable (with 3P adapter)
parallel I/O connector: 1
User’s manual, test result sheets: 1 set
Usage conditions 0℃ – 40℃, power supply voltage: 100 V 240 V ± 10% (Shipping configuration setting)
External dimensions (Not including carrying handle) W = 364mm H = 517mm D = 360mm

External dimensions