This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.

ITEMLIST
Products Infomation

ATSENSE - Flow Meter - FM-500
[Contact Us]
The FM-500 is an advanced engine-mounted fuel meter designed to deliver accurate net fuel consumption readings by measuring flow in both supply and return lines. For engines with return systems, it uses dual detectors to capture fuel flow data. This enables precise measurement of both instantaneous and total fuel usage. Developed by ATSENSE INC., located in Tokyo, it ensures reliable performance through innovative dual-sensor technology.
Feature
・Supports two flow detectors for both the supply and return
・Measurement of fuel consumption: supply flow rate – return flow rate
・High-speed response Analog output updated every 1ms
・Two-channel analog output Instantaneous consumption and total consumption
・USB data output Display data can be output as USB data simultaneously with analog output
・A single pulse conveniently output every 1mL for overall management of total flow
・Insulation Each input and output part is provided with isolation
・Erroneous operation is prevented by a key lock function
・Temperature and pressure measurement (option)
・Added function for temperature-corrected volume flow rate
Introduction to Functions
【Measurement of the return】
For an engine with return, the fuel returning to the supply side will present difficulties for the measurement. It is best in this case to do the detection with a second detector installed on the return side.
Using the signals from the two detectors on the supply and the return as inputs, the FM-500 measures both the instantaneous consumption and the total consumption.
FM-500 Flow meter for two inputs
Display | |
Display method | 6-digit x 2 rows, character height approx. 14.2mm 7-segment LCD, green |
Range of display and measurement | |
Instantaneous flow and average flow of total time |
-99.999 to 999.999 L/h |
Total flow | -99999.999 to 999999.999 L |
Pressure | ±100kPa (attained pressure) |
Temperature | -20.0 to 80.0℃ |
Display refresh time | 0.1s |
Accuracy | Excludes the accuracy of the sensor |
Instantaneous flow and average flow of total time |
±100ppm (RD) ±2digit |
Total flow | ±1digit |
Pressure | ±0.5kPa |
Temperature | ±0.5℃ |
Signal input | |
Number of inputs | 2 inputs (R03-R8F: Tajimi Electronics Co., Ltd.), R03-PB8M plug |
Supply voltage for the sensor | 12V, both supply and return 100mA (max) |
Flow sensor | |
Output signal format | A/B phase / UP/DOWN |
Trigger level | 2.5V (74HC365 equivalent) |
Voltage resistance | -5V to +15V |
Pull-up | Pulled up to 5V by 4.7kΩ |
Input frequency | 15kHz |
Temperature sensor | |
Element | Pt100, 3-wire system, excitation current 0.2mA |
Pressure sensor | |
Input range | 1-5V (3V = 0kPa) |
Contact input | |
Number of inputs | 2 inputs (7.62mm-pitch terminal block) |
Functions | Total measurement input and marker input when shorted to COM terminal |
Trigger level | 2.5V (74HC14 equivalent) |
Voltage resistance | -5V to +15V |
Pull-up | Pulled up to 5V by 4.7kΩ |
Pulse output | |
Number of outputs | 1 output (7.62mm-pitch terminal block) |
Output items | 1P/mL: single pulse output for each 1mL increase in total flow |
Output voltage | 0-5V (0.5V (max) voltage drop at 1kΩ load) |
Analog output | |
Number of outputs | 5 outputs (BNC receptacle), instantaneous consumption and total consumption |
Refresh time | 0.1s (switching at 1ms for instantaneous consumption only) |
Output range | ±10V (using D/A 16-bit) |
Accuracy | ±0.1% (FS) |
Allowable load resistance | 10kΩ |
Communication | |
Interface | USB mini-B (output of display values, and receiving at the PC side using general-purpose software) |
General specification | |
Insulation | Power supply input/sensor signal input and contact input/analog output and pulse output |
Power supply input voltage (power) | DC9V to 32V, 10W (max) |
External dimensions (mm) | W: 184 H: 70 D: 196 (excluding projecting parts) |
Weight | Approx. 810g |
Operating temperature and humidity range |
0 to 40℃/85%RH or less (with no condensation) |
Accessories | 5m power cable (1), and sensor connectors (2) |