This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.

ITEMLIST
Products Infomation

Handheld particle counters for accurate measuring and monitoring
Kanomax’s innovative Handheld Particle Counters (Models 3888 & 3889) set the standard of instrumentation in cleanroom / operating room monitoring and verification, filter testing, IAQ investigation, data center cleaning and many more highly specialized and highly sensitive applications. These beautifully engineered devices boast some of the most highly advanced technology in the industry. A combination of accuracy and practicality, these devices count airborne particles with high quality performance.
Since their inception, Kanomax particle counters have been the trusted choice in measuring instruments of by an ever-growing group of leaders in pharmaceutical, semiconductor, food and beverage industries. The 3888 and 3889 models perform with the same reliability and accuracy as their predecessors, but now have features that make them easier to use and that can gather measurements significantly faster.
They have very large, full color 4.3 inch LCD touchscreens, which can actually display data graphically for speedy interpretation. Despite the fact that measurements can very easily be taken directly from the instrument, these new models also include the easiest options for transferring data to PCs and other devices. There is an onboard USB input, which will accept a USB drive and download any number of the possible 10,000 data logs and transferred into a Windows 10 application.
The Kanomax Handheld Particle Counters meet ISO 21501-4 Standard and are fully RoHS compliant.
Handheld Particle Counters
Model:
Models 3888/3889
Maker:
KANOMAX
Features
Advanced Measurement Technology
Both models count airborne particles with exceptional accuracy and performance, featuring advanced technology that makes them standout instruments in their field.
User Interface
They are equipped with a large, full-color 4.3-inch LCD touchscreen, providing a user-friendly interface for easy data interaction and rapid interpretation of measurements.
Data Transfer and Management
These particle counters facilitate effortless data transfer to PCs and other devices, featuring an onboard USB input for downloading data logs (up to 10,000 records).
Compliance with Standards
The models comply with the ISO 21501-4 Standard and are RoHS compliant, ensuring they meet global standards for particle counting and environmental safety.
Model-Specific Features
Model 3888 displays 3 particle sizes (0.3, 0.5, 5.0μm), while Model 3889 displays 6 particle sizes (0.3, 0.5, 1.0, 3.0, 5.0, 10.0μm). Both models offer a minimum sensitivity of 0.3 microns and can store up to 10,000 measurements. They are compact, lightweight, and ergonomically designed for comfort during extended use.
Applications
These models are used for monitoring and verifying cleanrooms (ISO 14644: 2015 classification), monitoring surgical rooms, filter testing, IAQ investigation, and data center cleaning.
Technical Specifications
These models have a flow rate of 0.1CFM (2.83L/min), an accuracy of ±5% (compliant with JIS B9921 and ISO21501-4), a maximum measurable concentration of 10% for 2,000,000 particles/CF, and various other specd confirming their compliance and efficiency in particle counting.
Specifications
Main Unit Specifications | |||
---|---|---|---|
Product | Handheld Particle Counter | ||
Model | 3888 | 3889 | |
Particle Measurement | Sizes | 3 channels (0.3, 0.5, 5.0μm) | 6 channels (0.3, 0.5, 1.0, 3.0, 5.0, 10.0μm) |
Flow Rate | 0.1CFM (2.83L/min)
Accuracy: ±5% (Compliant with JIS B9921 and ISO21501-4) |
||
Maximum Measurable Concentration | 10% for 2,000,000 particles/CF
(compliant with JIS B9921 and ISO21501-4) |
||
Counting Efficiency | 50±20% (for PSL particles near the minimum measurable size)
100±10% (for PSL particles of 1.5 to 2 times as large as the minimum (Compliant with JIS B9921 and ISO21501-4) |
||
False Count | ≦1 particle/5 minutes
(Compliant with JIS B9921 and ISO21501-4) |
||
Size Resolution | ≦15% (for PSL particles near 0.3μm)
(Compliant with JIS B9921 and ISO21501-4) |
||
Display | 4.3 inch color LCD, Resistive touch panel | ||
Communication | Standard | USB (Host: for printer and USB flash memory, Device: for PC) | |
Stand (option) | RS485, Ethernet, and WiFi connection points in the base/stand | ||
Recording Media | Media | Internal memory | |
Number and Format | Up to 10,000 records in CSV format | ||
Language | English, Japanese, Chinese, and Spanish | ||
Power | Internal Supply | Li-ion rechargeable battery (Replaceable) | |
External Supply | AC adapter – Input 100 to 240 V | ||
Continuous Operation Time | 5 hours (Single measurement) | ||
Operating Environment |
Main Body | From 10 to 40 ℃, from 0 to 85%RH (With no condensation) | |
Probe | From 0 to 50 ℃, from 2 to 98%RH (With no condensation) | ||
Dimension | W213 X H100 X D55 mm | ||
Weight | 650g | ||
Hydrothermal Measurement (Available as an option for the 6- channel portable particle counter) |
Measurement Range | (Temperature) From 0 to 50℃
(Humidity) From 2.0 to 98.0%RH |
|
Display Resolution | (Temperature) 0.1℃
(Humidity) 0.1% |
||
Humidity Accuracy | ±3.0% (30 to 85%RH), ±5% (other range of humidity) | ||
Temperature Accuracy | ±0.5℃ | ||
Response Time | Approximately 60 seconds or less (90% response) |