global
Send us a message
product

ITEMLIST

Products Infomation

ADCMT DC Voltage Current Source Monitor 6253 / 6254

DC Voltage Current Source Monitor

Contact Us For Free Quote
 

The DC voltage current source/monitor 6253/6254 is suitable as power source for evaluation of semiconductors and other electronic components and as power source of characteristic systems. In addition to high accuracy, high stability and low noise of our strengths, the 6253/6254 realizes higher precision features such as 5½ digit source resolution, 6½-digit measurement resolution and basis accuracy of ±0.02 %.

Furthermore, the 6253/6254 can support various applications by the sweep functions, high-speed pulse source function with a minimum pulse width of 25 μs, the variable integration function and the newly adopted variable slew rate function.

DC Voltage Current Source/Monitor

Model:

6253 / 6254

Maker:

ADCMT

Download Materials

Features

• Source and measurement range

6253 Voltage: 0 to ±110 V, Current: 0 to ±2 A
6254 Voltage: 0 to ±20 V, Current: 0 to ±20 A

• 5½ source digits and 6½ measurement digits (measurement resolution: 100 nV/1 pA)

• High-speed pulse source with the minimum pulse width of 25 μs

• Variable slew rate function: 10.0 V/s to 99.9 kV/s

• Variable integration function

• Sink-enabled bipolar output

• Voltage and current parallel measurement by twin A/D converters

Applications

DC/DC converter evaluation

The following functions are effective for DC/DC converter characteristic test.

• The input current and the output voltage or current can be measured simultaneously by synchronizing two units.

• ±Source and ±sink by bipolar output.

 

The 6253/6254 allows 0 V sink which is unavailable by normal electronic load.

 

6253 DC/DC Converter Evaluation

 


Protection of overshoot at capacitive load

When a capacitive load is connected, overshoot occurs at the rising or falling edge and the DUT may be damaged or the output becomes unstable. Using the variable slew rate (SR) function will alleviate voltage waveforms and protect overshoot.

 

6253 Protection of Overshoot at Capacitive Load

 


Diode VF temperature dependence evaluation

Pulse current application is effective for power diode characteristic test to avoid the influence of self-heating.

 

By using the current pulse sweep function and voltage measurement in synchronization with pulses, precise VF characteristic test is available even with large current.

 

6253 Diode VF Temperature Dependence Evaluation

 


Average current measurement 【Variable Integration Function】

The integration time can be set arbitrarily from 100 μs to 1000 ms, making it easy to measure the average current consumption of cell phones, LCDs, and other devices.

 

As any integration time of the AD converter itself can be set and analog integration is adopted, there are no omissions in waveforms differently from digital integration, resulting in precise average measurement.

 

6253 Average Current Measurement

 

ADCMT DC Voltage Current Source/Monitor 6253

Specifications

 

Model 6253 6254
Number of digits for
generation
Output method Bipolar
Maximum output

(top)

Minimum resolution (bottom)

Voltage ±110 V/0.5 A ±20 V/7 A
5 μV
Current ±2 A/32 V ±20 A/7 V
50 pA 5 pA
Number of digits for

measurement

Accuracy (typical range) 0.02 %
Minimum

measurement

resolution

Voltage 100 nV
Current 1 pA 100 pA
Pulse application /

Measurement

Available
Minimum pulse width 25 μs
Interface USB/ GPIB/ LAN (option)/ RS232 (option)

Get Adobe' Reader'

To view the pdf files, the Adobe Reader from Adobe Systems is required.