This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.
ITEMLIST
Products Infomation
Semiconductor Curve Tracer CS-5000 Series
5,000V in the same size as the CS-3000 series
CS-5400/CS-5300/CS-5200 are the same size as CS-3300/CS-3200
CS-5100 is the same size as CS-3100
Semiconductor Curve Tracer
Model:
CS-5000 Series
Maker:
GRAPHTEC IWATSU Test Instruments Co.,Ltd.
Features
- Maximum peak voltage 5,000V (high voltage mode)
- Maximum peak current 1,500A (CS-5400 high current mode)
- All models equipped with LEAKAGE mode (1pA cursor resolution)
- USB port to save screen data and setup data
- Equipped with LAN interface for remote control
Line up
■ Curve Tracer CS-5400
- Max. Peak Voltage : 5,000V(HV mode)
- Max. Peak Current : 1,500A(HC mode)
■ Curve Tracer CS-5300
- Max. Peak Voltage : 5,000V(HV mode)
- Max. Peak Current : 1,000A(HC mode)
■ Curve Tracer CS-5200
- Max. Peak Voltage : 5,000V(HV mode)
- Max. Peak Current : 400A(HC mode)
■ Curve Tracer CS-5100
- Max. Peak Voltage : 5,000V(HV mode)
※ HC mode not available
Specifications
Collector supply HV mode Specifications
| Model | CS-5000 Series common | ||
|---|---|---|---|
| Mode/polarity | Full-wave rectification/ + – 、DC/ + – 、LEAKAGE/ + – 、AC | ||
| Maximum peak voltage | 5kV | 300V | 30V |
| Maximum peak current (maximum peak pulse current *) |
25mA (25mA*) |
750mA (1.5A*) |
7.5A (15A*) |
| Maximum peak power | At 5kV:320mW / 3.2W / 32W At 300V:120mW / 1.2W / 120W / 390W |
||
| Horizontal axis range | 50mV~500V/div | ||
* The maximum instantaneous current value that can be generated (varies depending on the target to be measured and conditions) measurement should be less than or equal to the maximum peak current value.
Collector supply HC mode Specifications
| Model | CS-5100 | CS-5200 | CS-5300 | CS-5400 | |||||
|---|---|---|---|---|---|---|---|---|---|
| Mode/polarity | HC mode Not mounted | Pulse/ + – | |||||||
| Maximum peak voltage | 40V | 30V | |||||||
| Maximum peak current | 400A | 40A | 1,000A | 400A | 40A | 1,500A | 600A | 60A | |
| Maximum peak power | 4kW | 400W | 10kW | 4kW | 400W | 12kW | 4.5kW | 400W | |
| Pulse width | Pulse width is variable at 50µs to 400µs (10µs resolution) | ||||||||
| Measurement point | The measurement point can be specified (10µs resolution) | ||||||||
| Vertical axis range | 100mA~50A/div | 100mA~100A/div | 100mA~200A/div | ||||||
| Test fixture | CS-303 | CS-304 | |||||||
Common Specifications
| Contents | |||
|---|---|---|---|
| Loop correction | Hardware | Correction of floating capacitance between collector supply and ground | |
| Software | Pseudo-looping through software optimization | ||
| Step generator | Offset | Setting range | S-10 ~ +10 x STEP AMPLITUDE Setting |
| Resolution | 1% of STEP AMPLITUDE setting | ||
| Current mode | Range | 50nA to 200mA per step, 21 steps 1-2-5 switching | |
| Maximum current | 2A | ||
| Maximum voltage | 10V or more | ||
| Voltage mode | Range | 50mV to 2V per step, 6 steps 1-2-5 switching | |
| Maximum current | 500mA~(~8V), 200mA~(~15V), 10mA~(~40V) | ||
| Maximum voltage | ±40V | ||
| Step rate | 2x 50Hz or 60Hz (1x in AC mode), pulse interval in HC mode | ||
| Pulse step | Pulse width | 50µs~400µs (10µs step) When in HC mode, the pulse width is 100µs wide before and after the pulse of the collector supply |
|
| Number of steps | 0~20 Step | ||
| AUX output | Range | OFF, -40V~+40V (Variable in 100mV steps) | |
| Measurement mode | REPEAT、STOP/SINGLE、SWEEP | ||
| Vertical axis (full scale 10div) |
Collector current | Range | HV mode:1µA/div to 2A/div, 20 steps 1-2-5 switching (HC mode is not included) |
| Accuracy | 2% + 0.05 x VERT/div of readout To this, add the loop correction error for the following maximum peak voltages 0.5µA(30V), 1µA(300V), 6µA(3kV), 12µA(5kV) Specify at least 10% of the maximum peak voltage for 30V, 300V, and 3kV, and at least 30% for 5kV |
||
| Emitter current (LEAKAGE) |
Range | 1nA/div~2mA/div, 20 step 1-2-5 switching (Collector supply mode is LEAKAGE) |
|
| Accuracy | 2% + 0.05 x VERT/div + 1nA or less of readout | ||
| Horizontal axis (full scale 10div) |
Collector voltage |
Range | HC mode:50mV/div to 5V/div, 7 steps 1-2-5 switching (HV mode is not included) |
| Accuracy | 2% + 0.05 x HORIZ/div or less of readout | ||
| Base/ emitter voltage |
Range | 50mV/div~5V/div, 7 step 1-2-5 switching | |
| Accuracy | 2% + 0.05 x HORIZ/div or less of readout | ||
| Display | Display | 8.4-inch TFT LCD | |
| Number of data points | 1,000 points/trace (AC, full wave rectification) 20~1,000 points/trace (SWEEP mode) |
||
| Trace display | Point interpolation display, dot display | ||
| Averaging | OFF, 2 to 255 times | ||
| Persistence | OFF, SHORT, LONG, Infinite length | ||
| Save internal waveform (REF) | 4-Panel | ||
| Cursor | DOT | Vert, Horiz, β or gm | |
| LINE | Vert, Horiz, 1/grad, intercept | ||
| FREE | Vert, Horiz, β or gm | ||
| WINDOW | Vert, Horiz, β or gm in the WINDOW area | ||
| Data save/read | Memory | Setup:256 pcs, REF:4 screens | |
| External memory | USB1.1:Setup, waveform, screen hardcopy (BMP, TIFF, PNG) | ||
| Remote | Equipped with remote 10BASE-T/100BASE-TX 1 port via LAN connection | ||
| Power supply | AC100~240V 50/60Hz, Maximum operating power:500VA (400W), Standby:50VA (7W) | ||
| External dimensions(mm) Weight |
CS-5100 | Approx. 424W x 220H x 555D (excluding protrusions and accessories) approx. 28kg (excluding accessories) |
|
| CS-5200, 5300, 5400 | Approx. 424W x 354H x 555D (excluding protrusions and accessories) approx. 43kg (excluding accessories) |
||