MICROTEST Impedance Analyzer 6632 | NIHON DENKEI (MALAYSIA) SDN.BHD. | Official Website
CONTENTS CONTENTS

ITEMLIST

Products Infomation

MICROTEST Impedance Analyzer 6632

Analysis of Impedance of Passive Components/Materials at Different Frequencies.

The MICROTEST 6632 Impedance Analyzer offers a wide frequency range of 10Hz to 50MHz for measuring the impedance characteristics of components and materials. The measured parameters include |Z| (impedance), |Y| (admittance), θ (phase angle), X (reactance), R (series/parallel resistance), G (conductance), B (Susceptance), L (inductance), D (dissipation factor), Q (quality factor), DCR (DC resistance), C (capacitance), ESR (equivalent series resistance), ε (relative permittivity), and μr (relative permeability). It supports both Meter mode and Sweep Analysis mode, allowing measurement results to be displayed in numeric or graphical formats. The system can be equipped with the optional Equivalent Circuit Analysis feature for three-component/four-component impedance analysis. It supports LAN/USB Host/USB Device/GPIB/RS-232 communication interfaces, making it compatible with automation equipment or human-machine interfaces (PLC).

Impedance Analyzer

Model:

6632

Maker:

MICROTEST

Evaluation impedance characteristics of RFID / NFC /  automotive wireless of antennas 

 

Equivalent Circuit Analysis (Option)

It has seven different models, combine with different types of parameters ( R, L, C) help customers model the impedance versus frequency characteristics and self-resonant frequency (SRF).

 

Four Parameters Displayed Simultaneously
7.0”TFT, 800×480 color screen, and four parameters can displayed simultaneously.

Measure Inductor (DCR)

Inductor coil turns and winding materials will affect DCR value. The key parameters for inductor are Ls / Lp / Q / SRF / DCR.

 

Evaluation of DC bias voltage characteristics with semiconductor wafer or ceramic multilayer capacitors.

 

  • Built-in the Dielectric constant-εr

 

 

  • Built-in the Permeability-μr

 

 

Features

  • Frequency range: DC, 10Hz to 1/3/5/10/20/30MHz/50MHz
  • Basic accuracy up to ±0.08% (typical ±0.05% )
  • Automatic Level Control (ALC)
  • Output impedance 25Ω/100Ω, switchable
  • Support meter mode and list mode, sweep mode
  • Equivalent Circuit Analysis 
  • Built-in DC Bias voltage ±12V
  • Ultra-high measuring speed < 3ms
  • Open circuit/short circuit/load correction function
  • Comparator function
  • BIN classification function
  • Built-in the Permeability-μr
  • Built-in the Dielectric constant-εr

Application

The induction coil for wireless charging in a mobile phone. Due to the manufacturing process, the wiring density layout from the induction coil will affect the original capacitance. The size of the parallel equivalent capacitance in the coil is the key to determine whether the product can be successful in high frequency application. It can affect the receiving function/wireless charging of an antenna. With this element analysis, we can accurately determine the RLC value to ensure the quality of the induction coil. 

 

Model E is suitable for measuring and analyzing the resonators, piezoelectric element or transducer. The two-resonance frequency characteristic diagram presented above is suited for Model E.

 

Testing PC board  inductance coil.

Utilizing the 6632 Impedance Analyzer to Measure Component Impedance Characteristics

Functions and Specifications

Model 6632
Frequency Range 10Hz-50MHz  (6632-50)

10Hz-30MHz  (6632-30)

10Hz-20MHz  (6632-20)

10Hz-10MHz  (6632-10)

10Hz-5MHz    (6632-5)

10Hz-3MHz    (6632-3)

10Hz-1MHz    (6632-1)

Frequency Setting Continuity
Frequency Resolution ±0.01%
Frequency Output Accuracy 100mHz, 6-bit Frequency Input
Basic Accuracy typical ±0.05%
AC Drive Level Test Signal Voltage Level

10mV – 2Vrms(100Ω)

10mV – 1Vrms(25Ω)

Voltage Minimum Resolution

1mV

Accuracy

ALC ON:6% * Voltage ±2mV

ALC OFF:10% * Voltage ±2mV

Test Signal Current Level

200µA – 20mArms

Current Minimum Resolution

10µA

Accuracy

ALC ON:6% * Current ± 20µA

ALC OFF:10% * Current ± 20µA

DC Drive Level 1V (fixed)
Output Impedance 25Ω/100Ω (switchable)
Measurement Mode Meter mode, list mode
Measurement Parameters |Z|、|Y|、θ、X、R、G、B、L、D、Q、DCR、C、ε and μr
Test Time (Fastest) < 3mS

 

Parameters Range
|Z| 0.000 mΩ to 9999.99 MΩ
R, X ± 0.000 mΩ to 9999.99 MΩ
|Y| 0.00000 μS to 999.999 kS
G, B ± 0.00000 μS to 999.999 kS
θRAD ± 0.00000 to 3.14159
θDEG ± 0.000° to 180.000°
Cs, Cp ± 0.00000 pF to 9999.99 F
Ls, Lp ± 0.00 nH to 9999.99 kH
D 0.00000 to 9999.99
Q 0.00 to 9999.99
Δ ± 0.00% to 9999.99%
Rdc 0.00 mΩ to 99.9999 MΩ
εr’ εr” 0 to 100000
μr’ μr” 0 to 100000

 

Measurement Circuit Series/Parallel
Correction Open Circuit/Short Circuit/Load correction
Cable Compensation 0 / 0.5 / 1 /2 (m)
BIN ABS、ΔABS、Δ%、OFF
Comparator 9
List Mode 50 groups of Multi-steps setting (Each group contains up to 15 steps)
Bulit-in Storge 100 sets LCR setting documents, 50 groups of list mode setting
USB Host Storge LCR setting documents, list mode setting document, BMP graphics, Sweep screen and test result data
Trigger Test Auto, manual, RS-232, GPIB, Handler
Power Supply Voltage:90-264Vac
Frequency:47-63Hz
Power consumption 30VA
Environment Temperature:10℃-40℃
Humidity:20-90%RH
Dimension 336x147x340 mm (W*H*D)
Weight 3 Kg
Interface RS-232、Handler、LAN、USB Host、USB Device
Display Color Screen, 7″ TFT (800*480)