MICROTEST Hipot Tester 7630

Supports AC withstand voltage / DC withstand voltage / insulation impedance measurement functions, capable of storing measurement data via USB, and supports PC connection software.

MICROTEST 7630 is a 3-in-1 test instrument that supports AC withstand voltage, DC withstand voltage, and insulation resistance testing functions. It has a maximum AC withstand voltage of 5000V with a leakage current of 30mA, a maximum DC withstand voltage of 6000V with a leakage current of 10mA, and a maximum insulation resistance of 12000MΩ. It supports PC connection software for remote programming and storage of test data. It can also be linked to a barcode scanner for barcode scanning, facilitating production traceability management.

Hipot Tester

Model:

7630

Maker:

MICROTEST

Safety Certification Testing Items

  • _AC Hi pot
  • _DC Hi pot
  • _Insulation Resistance
  • _Current Leakage

 

AC/DC Withstanding Voltage Test

Difference between AC and DC

AC Hi pot Test Most DUT have parasitic capacitance.

With AC test, might not able to charge the

parasitic capacitance. Therefore, this will

create constantly current go through.

DC Hi pot Test The parasitic capacitance will be charged under

DC hi-pot test. Therefore, the current will

reduce to almost 0. 

 

Advantages for AC Test

AC Hi pot Test AC test is same as our normal usage electricity.

Also, AC is able to do positive and negative

polarity test. 

 

AC test won’t charge the parasitic capacitance on the DUT, so it won’t create instantaneous peak

current. Do not need to discharge the DUT after

testing.

DC Hi pot Test The parasitic capacitance on DUT has been

charged under DC test, so the rest leakage

current is the real current from DUT.

 

Disadvantage of AC/DC Hi-pot Test

AC Hi pot Test If the DUT has huge parasitic capacitance, the current we detect will bigger then the actual current.

 

To fulfill the necessary current for the capacitance of the DUT, the output current is much more than DC test. This is danger for the operator.

DC Hi pot Test Testing Voltage has to increase from 0.

 

The higher parasitic capacitance has, the more

time it needs for DC voltage to increase. The voltage increase slower when it goes up. However, if

the charging current is too high, it may cause the misjudgment for the instrument.

 

The discharge process is necessary after DC

hi-pot test. 

 

DC test can only do the single polarity test. 

 

Hi-pot Test Time

The testing voltage is refer to the condition of DUT while doing Hi-pot test. If the voltage setting is too low, the insulation material defect might not be detected. If the voltage setting is too high and the testing time is too long, it may cause the permanent damage for the material.

Normally setting for the safety standard.

Formula  →Testing Voltage= DUT recommend voltage × 2+1000V 

EX : DUT standard :240V    →Setting test voltage:1480V

 

Common way to safe time for testing

Testing time for most products is 60s. If there are more than one test on one item, it will cost a lot of time. This will lower the efficiency og the product line.

Therefore, the product line usually shorter the testing time to 1~2 sec. and higher 10%~20% testing voltage.

Features

  • Hi pot ACW 5KV/30mA
  • Hi pot DCW 6KV/10mA
  • Insulation 1KV/12000MΩ
  • Arcing detection
  • Measuring Time (0.1 ~ 999s)
  • Adjustable ramp time (0.1 ~ 10s)
  • Safety Switch(INTER LOCK)
  • Option PC Link Software
  • Support AC 5000V, DC 6000V, Insulation Resistance 12GΩ

Application

What is Insulation Resistance Test
VDE and TUV safety standard request insulation resistance test for some specific product before hi-pot test. The reason is to make sure the insulation will not be damaged while doing hi-pot testing. Insulation resistance test is the key item to evaluate the material.

 

Hi-pot test is judge by leakage current, but IR test is judge by resistance, and usually it’s higher than 1 MΩ. It tests the DUT’s resistance between both terminal.

 

Why Do We Need To Do IR Test
Confirm the performance of the structure of the insulation.
The insulation structure should have higher insulation resistance rate by using better material.

 

For the insulation material, the protection against electric shock of the switch must be guaranteed.

Confirm the insulation structure can pass the reliability test(heat durability, insulation)

 

The importance of insulation resistance.

Before assemble the product, IQC has to do the IR test for every components, such as switches, transformers, resistors, capacitors, inductors, PCB, wire or cable to make sure the quality.

 

Leakage Current Test
→Measuring Tiny AC Leakage Current

DUT has tiny parasitic capacitance. This characteristic will create leakage current by resistance. Capacitor is a resistance component under AC voltage. This may cause the test result include the leakage current inside the instrument. If the DUT has low leakage current, we must consider capacitance , frequency, and the voltage we put.

 

Arc Detection
The surface or interior of the insulation material might discharge under high voltage test. Because of the lost of insulation performance, DUT may have temporary of discontinuity discharge.

Because that temporary discharge can’t be judge by leakage current, so Hi-pot tester will judge the defect product by the changing rate of testing voltage or leakage current.

 

Arc test is a geometric not a constant test. Also, the detection will be affect by the wire’s impedance or capacitance between the detectors. The rate of change of the current will be affect in the Arc detection.

Transformer safety regulation testing includes withstand voltage, insulation impedance, and leakage current measurements.

Specifications

Model 7630
Test Channel Single channel
AC Output Voltage 5000V
DC Output Voltage 6000V
Insulation Resistance 12000MΩ

 

AC Hipot
Output Voltage 10V-5000V
Voltage Resolution 1V
Voltage Accuracy ±(3% of setting+5V)
Rated Current (Max.) 30mA
Upper Leakage Current 0.001-31mA
Leakage Current Resolution 0.001mA
Leakage Current Accuracy ±(3% of reading +10uA)
Arcing Detection Detect gears from 0-20

(1 is the most sensitive)

DC Hipot
Output Voltage 10V-6000V
Voltage Resolution 1V
Voltage Accuracy ±(3% of setting+5V)
Rated Current (Max.) 10mA
Upper Leakage Current 0.001-11mA
Leakage Current Resolution ±(3% of reading +10uA)
Leakage Current Accuracy 0.001mA
Arcing Detection Detect gears from 0-10

(1 is the most sensitive)

Insulation Resistance
Output Voltage 10V-1000Vdc
Voltage Resolution 1V
Voltage Accuracy ±(3% of setting+5V)
Insulation Resistance Range 1-12000MΩ
IR Resolution 1MΩ
IR Accuracy 100-500V, 1-1000MΩ

(±5% of reading+1MΩ)

500-1000V, 2-12000MΩ

(±5% of reading +10MΩ)

Measuring Time 0.1-999s
Ramp Time 0.1-10s
Measuring Time Resolution 0.1s

 

PLC Remote Control Test、About
PLC Output Signal PASS、FAIL、HV Output、Testing
Safety Switch When testing, you need to short-circuit the INTER LOCK on the rear of the instrument to output the test voltage
List Mode Each group contains up to 32 steps
Built-in Storage 15sets setting files
USB Host Storage Storage setting files and update firmware
Power Supply Voltage:90Vac-264Vac
Frequency:50/60Hz ±5%
Power consumption 300VA
Environment Temperature:5℃-40℃
Humidity:0-80%RH
Dimension 365x145x430 mm (W*H*D)
Weight 14 Kg
Interface RS-232、Remote、USB Host、USB Device、EXT. I/O
Display 4.3″ TFT LCD (480*272 )