This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.


ITEMLIST
Products Infomation

HIOKI IMPEDANCE ANALYZER IM7587
Hioki LCR Meters and Impedance Analyzers range from 1 mHz to 3 GHz devices to suit a wide range of applications in the testing of electronic components. The IM7587 Impedance Analyzer offers a top measurement time of 0.5 ms over a 1 MHz to 3 GHz frequency range and superior stability, making it ideal for R&D as well as high volume production of ferrite chip beads and chip inductors.
1 MHz to 3 GHz Measurement Frequency Impedance Analyzer with 0.5 ms Test Speed and 0.07% Variability
IMPEDANCE ANALYZER
Model:
IM7587
Maker:
HIOKI
Key Features
- 1 MHz to 3 GHz testing source frequency
- Fastest test speed of 0.5 msec (Analog measurement time)
- 0.07% measured value variability (When measuring a 1 nH coil at 3 GHz)
- ±0.65% rdg. basic accuracy
- RF I-V method
- Half-rack size body and palm-sized test head
- Comprehensive contact check (via DCR testing, Hi-Z reject or waveform judgment)
- Make frequency sweeps, level sweeps and time interval measurements in Analyzer Mode
HIOKI IMPEDANCE ANALYZER IM7580 SERIES
Basic Spesification
Measurement modes | LCR mode, Analyzer mode (sweeps with measurement frequency and measurement level), Continuous measurement mode |
Measurement parameters | Z, Y, θ, Rs (ESR), Rp, X, G, B, Cs, Cp, Ls, Lp, D (tanδ), Q |
Measurable range | 100 mΩ to 5 kΩ |
Display range | Z: 0.00 m to 9.99999 GΩ / Rs, Rp, X: ± (0.00 m to 9.99999 GΩ) Ls, Lp: ± (0.00000 n to 9.99999 GH) / Q: ± (0.00 to 9999.99) θ: ± (0.000° to 180.000°), Cs, Cp: ± (0.00000 p to 9.99999 GF) D: ± (0.00000 to 9.99999), Y: (0.000 n to 9.99999 GS) G, B: ± (0.000 n to 9.99999 GS), Δ%: ± (0.000 % to 999.999 %) |
Basic accuracy | Z: ±0.65 % rdg. θ: ±0.38° |
Measurement frequency | 1 MHz to 3 GHz (100 kHz setting resolution) |
Measurement signal level | Power: -40.0 dBm to +1.0 dBm Voltage: 4 mV to 502 mVrms Current: 0.09 mA to 10.04 mArms |
Output impedance | 50 Ω (at 10 MHz) |
Display | 8.4-inch color TFT with touch screen |
Measurement speeds | FAST: 0.5 ms (Analog measurement time, typical value) |
Functions | Contact check, Comparator, BIN measurement (classification), Panel loading/saving, Memory function, Equivalent circuit analysis, Correlation compensation |
Interfaces | EXT I/O (Handler), USB communication, USB memory, LAN, RS-232C (optional), GP-IB (optional) |
Power supply | 100 to 240 V AC, 50/60 Hz, 70 VA max. |
Dimensions and mass | Main unit: 215 mm (8.46 in) W × 200 mm (7.87 in) H × 348 mm (13.70 in) D, 8.0 kg (282.2 oz) Test head: 90 mm (3.54 in) W × 64 mm (2.52 in) H × 24 mm (0.94 in) D, 300 g (10.58 oz) |
Included accessories | Test head ×1, Connection cable ×1, Instruction manual ×1, LCR application disc (Communications user manual) ×1, Power cord ×1 |