This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.
ITEMLIST
Products Infomation
Bench-type optical power meter suitable for R&D and production of optical disks
The 8250A is a bench-type optical power meter, which is most suitable for R&D and production of LDs, optical pickups, and drives for optical disks. It has 0.001 dB resolution and 5½-digit display and is also quipped with GPIB and USB interfaces as standard, allowing easy automated system establishment.
Nine kinds of optical sensors with various purpose (general-purpose, blue-violet, high-power, three-wavelength) and different shapes (thin-type, cylindrical-type) are available. The three-wavelength optical sensors 82314B, 82324B and 82314BW are capable of measuring 405 nm, 650 nm, 780 nm wavelength ranges. The blue-violet optical sensors 82312B and 82322B have realized a flat wavelength sensitivity characteristic and low dependency on an angle of incidence in the 405 nm range. The high-power optical sensors 82313B and 82323B enable high-power measurement of up to 200 mW.
Furthermore, the optical sensors for the TQ8210/TQ8215 series are also usable.
Optical Power Meter
Model:
8250A
Maker:
ADC CORPORATION
Features
- Wide lineup of sensors are available for each application
・Three-wavelength optical sensors for 405/650/780 nm
・Blue-violet sensors to measure 405 nm lasers
・High-power sensors to measure high-power lasers
・Low-priced general-purpose sensors
・Both thin types and cylindrical types - Optical power calibration with wavelengths of 405, 650 and 780 nm
- 5½-digit display and 0.001 dB resolution
- GPIB and USB interfaces
Specifications
| Display resolution |
0.1 pW (display in W), 0.001 dB (display in dBm) |
||
|
Accuracy (To be added to eash sensor accuracy) Conditoion: within 24 hours after offset zero execution, unit: W, 5½-digit display, CF=1 |
20 nW range |
±(0.55 %+2000 digits) |
|
| 200 nW range | ±(0.15 %+200 digits) | ||
| 2 μW to 200 mW range | ±(0 %+70 digits) | ||
|
Diplay |
VFD |
||
| Range switching | Auto, manual or remote switching among 8 ranges | ||
| Measurement speed |
FAST: 10 readings/s, MED: 5 readings/s, SLOW: 2 readings/s |
||
| Wavelength sensitivity correction |
Automatic correction of sensor wavelength sensitivity by wavelength setting (in 1nm step) |
||
| Interface | Analog output, USB, GPIB | ||