This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.


項目清單
產品資訊

用於監控電力消耗、能源管理、效能評估,以及確定電器或設備的運行效率
MICROTEST功率電錶7120應用於AC/DC和DC/DC電源相關的參數測量,量測頻率範圍從15Hz-100kHz,支援最高50階諧波分析,滿足Charger、Adapter、LED Driver、Inverter與DC/DC Converter等產品測試。
功率電錶(Power Meter)
Model:
7120
Maker:
MICROTEST
功率電錶應用於照明/LED、變頻電機、家電產品與太陽能光電等測量電源功率相關參數。
7120支援50階諧波分析功能
在電力系統諧波指的是電壓或電流的頻率為系統基頻的倍數,諧波通常都是因為負載或非線性元件所產生的(如磁飽和的磁性元件),關鍵是電網中的諧波是電力品質中最常見的問題,它會促使設備與導線額外的發燙,亦釀成馬達的轉矩脈動或是閘流體整流模組誤觸發。為滿足電源功耗更精密的測量,7120支援50階諧波分析能力,可測量高達100kHz的元件,達到高階諧波分析功能。
功能特點
- AC/DC兩用安培/瓦特小時計
- 寬廣量程範圍0.001W~16KW
- 最高50階諧波分析功能
- 附PC Link軟體
- 具波峰因素比顯示
- 1000組量測數據保存空間
- 電流波峰因數最高CF9
- 自動換檔速度快
- 待機功率解析度達 0.001W
MICROTEST功率電錶功能
規格
Model | 7110 |
Frequency Range | DC, 15Hz~10KHz |
Harmonic (option) | ● / – |
Model | 7120 |
Frequency Range | DC, 15Hz~100KHz |
Harmonic (option) | ● / – |
Parameters | Range |
Vrms | 0.1V ~ 600V |
Vdc | 0.1V ~ 600V |
Irms | 0.1mA ~ 20A |
Idc | 0.1mA ~ 20A |
W | 0.01W ~16KW |
PF | 0.000 ~ ±1.000 |
Deg | -180°~ +180° |
THD | 0.01% ~ 99.99% |
Hz | 15Hz ~100kHz |
Measurement Range | |
Current (fixed/Auto) | 0.01A、0.03A、0.1A、0.3A、1A、3A、10A
20A |
Voltage (fixed/Auto) | 10V、30V、100V、300V、600V |
Harmonics SPEC | |
Analysis base | To achieved stable fundamental frequency
analysis by voltage or current (non-inverter) |
Frequency Range 45Hz | 45Hz – 440Hz |
FFT Data Length | 1024 |
FFT Data Format | 32 bits |
Measurement
Projects
|
1-50 THD, 1-50 level voltage and current
V [n], A [n] 1-50 level voltage and current distortion percentage V [n%], A [n%] 1-50 level watts W [n] 1-50 level watts distortion percentage Watt W [n%] 1-50 level voltage and current angle DEG [n] Vrms, Irms, Watt, PF |
Model | 7110
7120 |
Frequency Measurement Mode | To achieved stable base frequency measurement (variation less than 1%) by voltage or current (non-inverter) |
Frequency Range | Model 7110
DC, 15Hz~10KHz |
Model 7120
DC, 15Hz~100KHz |
|
Data Length | Dual 4096×16 RAM for voltage & current |
ADC Resolution | 16 bits |
Sampling Rate | AC 50Hz/60Hz basic sampling rate 100
KSPS /120 KSPS |
Arithmetic Precision | Watt/VRMS/IRMS/MEAN/PF/Deg/Line filter 32bits |
Frequency Filter | 500Hz cut off, digital chip filter based on
25MHz |
Signal Filter | 500Hz-3db digital filter based on
Butterworth 50Hz-0.03% reading, 60Hz-0.05% reading |
Frequency
Acquisition Mode |
Voltage / current 100MHz baseband
digital dynamic meter chip |
Phase Lead Detection | Subject to the current, analog/digital
hybrid detecting (error less than 5 degrees) |
RMS/MEAN Mode Voltage & Current Accuracy (23℃ ±5℃ ) | |
>=15 Hz, f ≦ 45Hz | ±(0.1% of reading + 0.4% of range) |
>45 Hz, ≦ 66 Hz | ±(0.1% of reading + 0.1 % of range) |
>66 Hz, ≦ 1 kHz | ±(0.1% of reading + 0.2 % of range) |
>1 kHz, ≦ 10 kHz | ±(0.07*f % of reading + 0.3% of range) |
>10 kHz, ≦100 kHz | ±(0.5% of reading + 0.5% of range)
±[{0.04×(f–10)}% of reading] |
Funit is 1KHz
When the L-FILTER sets as ON: 45Hz~66Hz frequency range allowable error-0.03 %~-0.05 of reading |
|
When the AC is measured, if the fundamental frequency exceeds 200Hz, the F-Filter is required to be turned off in order to measure the most accurate
value
**When the frequency range is more than 10KHz, the 7120 starts to support |
DC Mode Voltage & Current Accuracy (23℃ ±5℃ ) | |
10V~600V | ±0.2% reading ± 0.2% of range |
0.01A ~20A | ±{(0.2)% of reading + 0.2 % of range} ±offset |
To add up the OFFSET errors of various files during measuring the DC
current |
Power (W) Accuracy (23℃ ±5℃ ) | |
AC power ranges (Auto or Manual)(40 ranges) range up to 16KW
Maximum Power (W) value is determined by the highest range of voltage profile |
|
DC ±0.2% reading ± 0.5% of range
15 Hz≦f 66 Hz<f≦1 kHz ±(0.2% of reading + 0.2 % of range) 1 kHz<f≦10 kHz ±(0.4% of reading + 0.3 % of range)± [{0.06×(f)}% of reading] 10 kHz<f≦100 kHz ±(0.5 % of reading + 0.5 % of range)±[{0.09×(f–10)}% of reading] |
|
Incidental Allowable Error Conditions | |
Signal Filter Error (AC) | Frequency between 45-66Hz: Add 0.3% of
reading. |
Frequency between45-66Hz: Add 1% of
reading beyond |
|
Accuracy Effect of the Phase Error of the Power | |
When the power factor PF is 0, the error range of Watt is Situation 1: for 45Hz < f, Add±1.0% of VA Situation 2: for 45Hz > f or f > 66Hz Add ±{(3.5 + 0.5×f )% of VA} for up to 100kHz as reference data The unit for frequency f is kHz. When the power factor is 0 < PF≦error range When 0 < PF 1 (θ: phase angle of the voltage and current) for 45Hz f 66Hz. Add ±power reading *{tan(θ)*(0.5)}% for f < 45Hz, f > 66Hz. Add ±power reading *{ tanθ*(0.5×f+0.2) }% |
|
Error within 12 months | Add ±(0.5% of reading) |
Model | Support Harmonics
7110-10k-HARM 7120-100k-HARM |
Don’t Support Harmonics
7110-10k 7120-100k |
|
F71201 TEST BOX | |
Built-in stroage | 6 Sets |
Power Supply | Voltage
100 ~ 240Vac |
Frequency
50/60Hz |
|
Power consumption | 300VA |
Environment | Temperature
23℃±5℃ |
Humidity
20%-80%RH |
|
Display | Seven-segment display |
Dimension (W*H*D) | 227x101x300 mm |
Weight | 1.85 Kg |
Interface | RS-232 (7110)、RS232 + GPIB (7120) |