Of a bench-top size that can fit on a personal desk, this small temperature and humidity chamber achieves high performance to fulfill your environmental testing needs.
The lineup including 12 models available in two sizes (22.5L and 64L) with a temperature range of -20, -40, or -60 °C to +150 °C for temperature tests or climate tests allow you to choose the performance best suited for your test applications.
In addition, the model with temperature change rate of 5K/min can meet requirement for fast cycling test from various industries.
Featuring a touch screen color display for easy operation, the controller carries the same function and design of the Platinous J chambers.

Compact Environmental Tester

Model:

SU-662

Maker:

ESPEC CORP.

Download Materials

Features

Compact enough for desktops
 
Stackable for enhanced space efficiency
Caster-equipped stands are easy to move and are available in low types (C stand/L stand), H stand for storing these low types underneath, and M stand that comes in the same height as commonly used office desks.
 
Instrumentation Interlock I/O Terminals
Chamber comes with instrumentation interlock I/O terminals as a standard, which allow the chamber to work in synchronization with measuring instruments.
Using an input terminal enables chamberspecified program patterns to be run or stopped depending on measuring instrument operation.
 
Three-way access
Chamber comes with a cp50mm cable port on the right side as a standard, but you can enlarge and or add more cable ports on the right, left and the top.

Line up

Model Power Supply Temperature & Humidity range Inside/Outside dimensions(mm)
SH-222 100V AC 50/60Hz
115V AC 60Hz NEC
220V AC 50/60Hz *
230V AC 50Hz *
-20 to +150℃
30 to 95%rh
W300×H300×D250
SH-242 -40 to +150℃
30 to 95%rh
SH-262 100V /200V AC 50/60Hz
220V AC 50/60Hz *
230V AC 50Hz *
200V AC 50/60Hz NEC
-60 to +150℃
30 to 95%rh
W300×H300×D250
SH-642 -40 to +150℃
30 to 95%rh
W400×H400×D400
SH-662 -60 to +150℃
30 to 95%rh
SH-242-5
Temperature rate of change:5℃/min
100V /200V AC 50/60Hz
220V AC 50/60Hz *
230V AC 50Hz *
-40 to +150℃
30 to 95%rh
W300×H300×D250

* Compliance with CE marking.

Specifications

SU−662
System Balanced Temperature Control system (BTC system)
Temp. performance*1

Temp. range −60 to +150°C (−76 to +302°F)
Temp. fluctuation ±0.3°C (−60 to +100°C) ±0.5°C (+100.1 to +150°C)
Temp. variation in space 2.5°C (−60 to +100°C) 4.0°C (+100.1 to +150°C)
Temp. rate of change Heat up rate 2.9°C/min
Pull down rate 1.7°C/min
Temp. extreme achievement time Heat up time From −60 to +150°C within 80 min.
Temp. extreme achievement time Pull down time From +20 to −60°C within 90 min.
Lowest attainable temp. −60°C
Construction
Heater Nichrome strip wire heater
Refrigeration unit System Mechanical cascade refrigeration system
Cooler Plate fin cooler
Refrigerator Hermetically sealed compressor, Air-cooled condenser, Expansion mechanism: Capillary tube system
Refrigerator capacity 400W [Unit 1: 400W × 1, unit 2: 400W × 1]
Refrigerant R-404A, R-23
Capacity 64 L
Chamber total load resistance 20 kg
Inside dimensions mm (inch) *2 W400×H400×D400 (W15.7 × H15.7 × D15.7)
Outside dimensions mm (inch) *2 W540×H665×D921 (W21.2 × H26.1 × D36.2)
Weight 123 kg
Utility requiremnets
Allowable ambient conditions +5 to +35°C (+41 to +95°F)
Power supply *3
100V AC 1φ50/60Hz 17.5 A
115V AC 1φ60Hz (NEC) —
200V AC 1φ50/60Hz *4 14.5 A
220V AC 1φ50/60Hz *5 14.0 A
230V AC 1φ50Hz *5 14.0 A
Noise level *6 Between 48 and 53 dB
Exhaust heat quantity 5040 kJ/h

*1 The performance values are based on IEC 60068-3-5:2001 for the temperature chamber.
Performance figures are given for a +23°C ambient temperature, 65%rh, rated power supply and no specimens inside the test area. However, the lowest attainable temperature is given for a max. ambient temperature of +30°C. Heat-up time is the achieved time from lowest temperature to highest temperature within temperature range.
*2 Excluding protrusions.
*3 At ambient temperature +23°C.
*4 200V AC available with or without NEC specifications. SU-242-5 not available with NEC specification.
*5 Compliance with CE marking.
*6 Measurements are to be taken in an anechoic room at a height of 1.2m from the floor and a distance of 1m from the chamber front panel
(ISO 1996-1:2003 _ A-weighted sound pressure level)