Condenser microphone has excellent features, in its frequency response, stability and so on. Therefore, it is mainly used for measurement.

Condenser Microphones

Model:

UC・NH_SERIES

Maker:

RION CO.,LTD.

Features

 

Condenser Microphones UC Series
Sound field and sound pressure models
In ordinary measurements, the placement point at the microphone has an impact on sound pressure at the measurement point at high frequencies. Sound field microphones are designed to flatten frequency characteristics by taking that impact into account in advance. They are used for ordinary measurements.
In contrast, sound pressure microphones are designed to have flat frequency response for the sound pressure applied to the diaphragm without taking the impact on the surrounding sound field into account. They are used for measurements in reverberant room or using couplers.

 

Preamplifiers NH Series
Preamplifier Overview
A preamplifier with high input impedance and low output impedance is required in order to faithfully transfer the voltage converted by the microphone to the subsequent amplifier.
In the preamplifier, the electrical circuit and 7-pin connector have a unified structure and various models are available to ensure the optimal combination for the type of microphone and diameter. A conversion adapter is used when using a microphone with a different diameter.

Microphones

 

Model UC-27 UC-34P  UC-35P
Suitable preamplifier NH-06A NH-34 set NH-35 set
Nominal diameter 1 inch
Frequency response Field Field Field
Frequency range (Hz) 5 to 12 500 10 to 12 500 10 to 12 500
Bias voltage (V) 200 200 0
Sensitivity level (dB re 1 V/Pa) *1 -26.5 105 0
Capacitance (pF) 54
Maximum input sound pressure level (dB)
(Linearity error ±0.3 dB)
152 96
A-weighted Inherent noise (dB) 12 2 4
Temperature coefficient (dB/°C) -0.005
Diaphragm Titanium alloy
Dimensions (mm) 23.8 (dia)×21.0 23.8 (dia)×131 23.8 (dia)×132.7

 

Model   UC-30  UC-31 UC-33P  UC-52  UC-59  UC-57
Suitable preamplifier NH-04A/05A/12A NH-04A/05A/12A NH-04A/05A/12A NH-17/17A/22A NH-17/17A/22A NH-17/17A/22A
Nominal diameter 1/2 inch
Frequency response Field Field Pressure Field Field Field
Frequency range (Hz) 10 to 20 000 10 to 35 000 10 to 20 000 20 to 8 000 10 to 20 000 10 to 16 000
Bias voltage (V) 200 200 200 0 0 0
Sensitivity level (dB re 1 V/Pa) *1 -25.5 -37 -38 -33 -27 -22
Capacitance (pF) 17 20 20 19 13 14
Maximum input sound pressure level (dB)
(Linearity error ±0.3 dB)
144 160*4 160 150 148 132*4
A-weighted Inherent noise (dB) 20 26 28 24 18 13
Temperature coefficient (dB/°C) -0.007 -0.007 -0.009 -0.008 <±0.35 dB (at 1 kHz)*3 <±0.45 dB (at 250 Hz)*3
Diaphragm Titanium alloy
Dimensions (mm) 13.2 (dia)×15.0 13.2 (dia)×13.2 13.2 (dia)×13.0 13.2 (dia)×12.0 13.2 (dia)×14.3 13.2 (dia)×13.5

 

Model  UC-29  UC-54
Suitable preamplifier NH-05A (UA-12 required) NH-17/17A/22A (UA-12 required)
Nominal diameter 1/4 inch
Frequency response Field Field
Frequency range (Hz) 20 to 100 000*2 20 to 100 000*2
Bias voltage (V) 200 0
Sensitivity level (dB re 1 V/Pa) *1 -47 -48
Capacitance (pF) 6 4
Maximum input sound pressure level (dB)
(Linearity error ±0.3 dB)
164*4 164
A-weighted Inherent noise (dB) 42 45
Temperature coefficient (dB/°C) -0.01 <±0.7 dB (at 250 Hz)*3
Diaphragm Titanium
Dimensions (mm) 7.0 (dia)×10.0 7.0 (dia)×10.0

*1: 1 kHz Representative value

*2: UC-29/54 frequency range is the one (measured on condition) that the grid at its top is taken off.

*3: -10 ˚C to +50 ˚C (+23 ˚C is reference point)

*4: Distortion rate 3 %

*5: Depend on connected instrument

Microphone with Preamplifiers & Preamplifiers

 

Microphone with Preamplifiers (TEDS compliant)

Model  UC-52T  UC-57T UC-59T
Microphones UC-52 UC-57 UC-59
Preamplifier NH-22AT NH-22AT NH-22AT
Nominal diameter 1/2 inch
Frequency response Field Field Field
Frequency range  (Hz) 20 to 8 000 10 to 16 000 10 to 20 000
Constant current drive 2 mA to 4 mA 2 mA to 4 mA 2 mA to 4 mA
A-weighted Inherent noise (dB) 24 13 18
Dimensions (mm) φ13.2×97 φ13.2×98.5 φ13.2×99.4
Cable type EC-90 series (BNC) EC-90 series (BNC) EC-90 series (BNC)

TEDS is a format to describe transducer information regulated by IEEE 1451.

Sensors will be calibrated automatically when connected to TEDS applicable instruments as it has the following recorded parameters inside

 

Preamplifiers

Model NH-06A NH-04A NH-12A NH-17 NH-17A   NH-22A NH-05A
Suitable microphones UC-27 UC-30/31/33P UC-30/31/33P UC-52 /54*1/ 57/59 UC-52 /54*1/ 57/59 UC-52/54*1/57/59
(constant current drive) 2 mA to 4 mA
UC-29*1 UC-30 /31 /33P
Nominal diameter 1 inch 1/2 inch, 1/4 inch*1 1/2 inch, 1/4 inch*1
Input impedance (GΩ) 3 3 3 3 3 6 10
Input capacitance (pF) 0.3 0.25 0.25 0.8 0.8 0.7 0.2
Frequency range (Hz) 5 to 100 000 10 to 100 000 10 to 100 000 10 to 100 000 10 to 100 000 10 to 100 000 10 to 100 000
Bias voltage (V) 200 200 200 0 0 0 200
Gain (dB) -0.1 (54 pF)
(UC-27)
-0.2 (17 pF)
(UC-30)
-0.2 (17 pF)
(UC-30)
-0.5 (13 pF)
(UC-59)
-0.5 (13 pF)
(UC-59)
-0.5 (13 pF)
(UC-59)
-0.5 (6pF)
(UC-29*1)
A-weighted Inherent noise (dB) 12 (UC-27) 20 (UC-30) 20 (UC-30) 18 (UC-59) 18 (UC-59) 18 (UC-59) 42 (UC-29)
Output impedance (Ω) <_ 100 <_ 100 <_ 100 <_ 300 <_ 300 Approx. 30 <_ 100
Cable type EC-04 series (7p) EC-04 series (7p) 1.5 m integrated (7p) 5 m integrated (7p) EC-04 series (7p) EC-90 series (BNC) EC-04 series (7p)

*1: UA-12 required