This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.
ITEMLIST
Products Infomation
Compiled with the test voltage -25 V to -1000 Vdc of the JIS C 1302-2002
The TOS7200 is an insulation resistance tester available for a wide range of various electric and electronic components, as well as electric and electronic equipment. The output voltage can be set at desired value in the range of – 25V to -1,000V with a resolution of 1V. (conforms with the output characteristics of the JIS C 1302-2002).
As it is fitted with a window comparator and timer function, the tester is capable of efficiently conducting insulation resistance tests based on various safety standards. In addition, this product is equipped with panel memory as standard feature, which can be recalled by remote control, SIGNAL I/O connector, and the RS-232C interface for easy automatic testing system construction.
Insulation Resistance Tester
Model:
TOS7200
Maker:
KIKUSUI ELECTRONICS CORPORATION
Features
- Output voltage setting range: -25 to -1000V
- Resistance measurement range: 0.01MΩ to 5000MΩ
- Equipped with a discharge function
- Equipped with a window comparator
- Hold function (holds the measured resistance value during the PASS period at the end of the test)
- Equipped with a timer function
- Rear output terminals
- Measured-value monitoring terminals
- Equipped with the panel memory (enabling 10 different settings to be stored)
- SIGNAL I/O, Remote control terminals
- RS-232C interface included as standard
Resistance Mmeasurement Range
| Test voltage | Resistance measurement range |
|---|---|
| 25V | 0.03MΩ to 250MΩ |
| 50V | 0.05MΩ to 500MΩ |
| 100V | 0.10MΩ to 1000MΩ |
| 125V | 0.13MΩ to 1250MΩ |
| 250V | 0.25MΩ to 2500MΩ |
| 500V | 0.50MΩ to 5000MΩ |
| 1000V | 1.00MΩ to 5000MΩ |
Specifications
Environment
| Installation location | Indoors and at altitudes up to 2000 m |
|---|---|
| Warranty range | Temperature 5 °C to 35 °C (41 °F to 95 °F) |
| Humidity 20 %rh to 80 %rh (no condensation) | |
| Operating range | Temperature 0 °C to 40 °C (32 °F to 104 °F) |
| Humidity 20 %rh to 80 %rh (no condensation) | |
| Storage range | Temperature -20 °C to 70 °C (-4 °F to 158 °F) |
| Humidity 90 %rh or less (no condensation) |
Input voltage range
| Nominal voltage range (allowable voltage range) | 100 V to 240 V AC (85 V to 250 V AC) |
|---|---|
| Power consumption at rated load | 30 VA maximum |
| Allowable frequency range | 47 Hz to 63 Hz |
| Insulation resistance | 30 MΩ or more (500 V DC) [AC LINE to chassis] |
| Withstanding voltage | 1390 V AC for 2 seconds, 10 mA or less [AC LINE to chassis] |
| Earth continuity | 25 A AC / 0.1 Ω or less |
Safety *1
Conforms to the requirements of the following standard.
EN 61010-1 (Class I *2 , Pollution degree 2 *3)
*1 Not applicable to custom order models.
*2 This instrument is a Class I equipment. Be sure to ground the protective conductor terminal of the instrument. The safety of the instrument is not guaranteed unless the instrument is grounded properly.
*3. Pollution is addition of foreign matter (solid, liquid or gaseous) that may produce a reduction of dielectric strength or surface resistivity. Pollution Degree 2 assumes that only non-conductive pollution will occur except for an occasional temporary conductivity caused by condensation
Dimensions (maximum) /Weight
215[8.46 inch] W × 66[2.60 inch] (85[3.35 inch]) H × 230[9.06 inch] (260[10.24 inch]) D mm / Approx. 2 kg (Approx.4.41 lbs)
Accessories
AC power cable: 1 pc. / TL08-TOS high-voltage test leadwires (1.5 m): 1 set / Setup guide: 1 pc. / Quick reference: English: 1pc., Japanese: 1pc. / Safety information: 1 pc. / CD-ROM: 1 pc.