This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.


ITEMLIST
Products Infomation

Instrument capable of accommodating test with and Without Cl2
This test instrument is capable of performing a gas corrosion test specified in the standard of ISO, IEC, JIS, etc.to evaluate the corrosion resistance of the electronic parts, plating, etc., and the storage stability of image.
Referring to IEC 60068-2-60(Environmental Testing -Part 2 : Test Ke : Flowing mixed gas corrosion test), a chlorine has a highly reactive , and once a test containing the chorine has been conducted, the chlorine shows strong interaction especially through reaction with hydrogen sulfide, there is a possibility of influencing test results. Moreover, it is extremely difficult to completely eliminate the chlorine from the sections that have been exposed to chlorine.
Gas corrosion Test Instrument
Model:
GT-100
Maker:
Suga Test Instruments Co., Ltd.
Specification
Gas Concentration Range |
SO2 | 0.1~200ppm (100~200,000ppb) |
H2S | 0.01~200ppm ( 10~200,000ppb) | |
NO2 | 0.1~ 20ppm (100~20,000ppb) | |
Cl2 | 0.01~ 20ppm ( 10~20,000ppb) | |
Temperature and Humidity Range |
Temperature | 20~65±1℃ |
Humidity | 65~95±3%rh at 20℃ | |
45~95±3%rh at 40℃ | ||
30~95±3%rh at 65℃ | ||
Effective Test Chamber Dimensions | Approx. Width 50 × Depth 50 × Height 40 cm | |
Main Unit External Dimensions | Approx. Width 130 × Depth 144 × Height 194 cm | |
Electrical Requirement | 3-phase 200V approx. 19A 50/60Hz |