RION Tri-axial Ground borne Vibration Meter VM-56 | NIHON DENKEI PHILIPPINES, Inc.
CONTENTS CONTENTS

ITEMLIST

Products Infomation

RION Tri-axial Ground borne Vibration Meter VM-56

Simultaneous PPV, VDV, Dominant Frequency & Displacement

The VM-56 is a ground borne vibration meter capable of simultaneously calculating the measurement quantities defined by DIN 45669-1, ISO 8041 and other national measurement standards. Like other Rion products, it is characterized by excellent build-quality and exceptional ease of use. It is suitable for a wide range of applications including attended measurements, unattended surveys and live-to-web monitoring.

Tri-axial Ground borne Vibration Meter

Model:

VM-56

Maker:

RION CO.,LTD.

AppIicabIe standards

  • DIN 45669-1 : 2010-09
    (Measurement of vibration immission —Part 1 : Vibration meters ー Requirements and tests)
    * Measurement range, measurement frequency range only
  • ISO 8041 : 2005, ISO 8041-1 : 2017 (Human response to vibration ー Measuring instrumentation

Tri-axial Ground borne Vibration Meter VM-56

Specifications

 

Applicable standards DIN 45669-1: 2010-09 (Frequency, Measurement range compliance), SBR Meten en beoordelen van trillingen, Deel A: Schade aan gebouwen 2010, Deel B: Hinder voor personen 2013, ISO 8041: 2005, ISO 8041-1: 2017, CE marking, WEEE directive
Measurement functions Tri-axial simultaneous measurement
Measurement values In accordance with DIN Peak particle velocity |v|max (PPV)
Dominant frequency fmg (D.F.)
Weighted vibration maximum value KBFmax
Maximum KBF value over 30-second cycle KBFT
In accordance with ISO Corrected acceleration effective value Acc.
Maximum transient vibration value MTVV
Vibration dose value VDV
Crest factor C.F.
In accordance with SBR Maximum weighted vibration value veff, max
Maximum veff over 30-second cycle  veff, max, 30
Others Displacement (0-p value) Disp.
Combined PPV for 3 axes PVS
Waveform recording (Option) Time waveform of acceleration signal a(t)
1/3 octave band analysis value (Option) Time-weighted time average, maximum acceleration
Tri-axial synthesis of band max overall Law
Measurement frequency range 0.5 Hz to 315 Hz
Frequency bandwidth limits For acceleration, velocity, and displacement signals, the following frequency range limits can be selected.
Lower limit: 0.5 Hz, 1 Hz, 4 Hz
Upper limit: 80 Hz, 100 Hz, 250 Hz, Sensor Dependent (LPF OFF)
Measurement range Measurement frequency setting is 1 to 80 Hz, defining the following range
Measurement range for VM-56 Vibration velocity: 0.03 to 100 mm/s
Weighted vibration amount: 0.02 to 100 mm/s (Reference 16 Hz)
Maximum absolute waveform value: 0.05 to 100 mm/s (Reference 16 Hz)
Vibration acceleration: 0.0003 to 10 m/s2
Displacement (0-p): 0.01 to 10 mm (0.5 to 4 Hz)
Measurement range compliant with SBR-Deel B
Vibration velocity: 0.02 to 100 mm/s (Frequency bandwidth 1 to 80 Hz)
Instrument noise Vibration acceleration 0.0001 m/s2 (Measurement frequency range 1 to 80 Hz)
Vibration velocity Max. 0.01 mm/s (Measurement frequency range 1 to 80 Hz)
Frequency correction No weighting (Common band filter for ISO and DIN / SBR band filter)
KB (DIN 45669-1 compliant)
Wb, Wd, Wm characteristics (ISO 8041 compliant)
Hv (SBR-B compliant)
Measurement range 2 switchable ranges, separate for 3 axes: 0.001 to 10 m/s2, 0.0001 to 1 m/s2
Dynamic range Max. 100 dB
Sampling frequency 2 kHz
Store modes 3 modes (Manual, Auto, Timer Auto), Data format: CSV
Manual Measurement results stored with measurement start time in one memory address Data stored in internal memory or on SD card (Internal memory: max. 1000 tri-axial data sets, SD card: dependent on card capacity)
Processed value store: PPV, Dominant Frequency (D.F.), KBFmax, MTVV, VDV, Crest Factor (C.F.),
Displacement (Disp.), PVS, Overload and Under Range Flags for each calculation cycle.
Auto Continuous storing of various types of processing results for each calculation cycle Data stored on SD card Store modes: Instantaneous store, calculation store, level trigger store
• Instantaneous store: Acc. rms data stored every 100 ms
• Processed value store: PPV, Dominant Frequency (D.F.), KBFT, MTVV, VDV, Crest Factor (C.F.), Displacement (Disp.), PVS, Overload and Under Range Flags for each calculation cycle.
• Calculation cycle: 1 s to 24 h
Timer Auto Processed values are continuously recorded for each store cycle at the set measurement start / stop time.
Sleep function (power save mode until measurement start) available Data stored on SD card Store modes: Instantaneous store, Calculation store Sleep function (power save mode until measurement start) available Data stored on SD card Store modes: Instantaneous store, Calculation store
• Instantaneous store: Acc. data stored every 100 ms
• Calculation store: Processing results for each calculation cycle
• Calculation cycle: 1 s to 24 h
Measurement time Max. 200 days (Auto store mode only, with 100 ms off)
Data recall Store data name, store data browse, time browse, waveform yes/no check
Setting memory Up to 5 sets of settings can be stored in internal memory and on SD card, for later recall
Startup with settings stored in a file on the SD card possible
Clock function Year/Month/Day/Hour/Minute/Second, Daily error ±1 s, 10 ppm
Display Backlit semi-transparent color TFT LCD, WQVGA resolution (400 x 240 dots)
Language: English only
Alarm indication Signal overload indication, signal underload indication
Signal output 2.5 dia. output jacks, 3 separate channels
AC output AC output: 1 Vrms (full-scale)
Frequency weighting for instantaneous value display and for
AC output can be set separately
Frequency range: 0.5 to 315 Hz
USB Mass storage class: SD card recognized as removable disk
Communication device (virtual COM port): Supports command based communication
RS-232C communications Using dedicated cable (I/O terminal)
Comparator output Open-collector output (using I/O port)
Max. applied voltage: 24 V
Max. drive current: 50 mA (with 24 V applied voltage)
Monitored Parameter: PPV (broad-band or user-definable PPV vs frequency function)*3
Power requirements IEC R6 [size AA] battery x 8 or external power supply
Battery life (23 ℃) Alkaline battery LR6 (AA): 24 h , Ni-MH secondary battery: 24 h
*Battery life will differ depending on settings.
AC adapter NC-98E
External power supply voltage 5 to 7 V (rated voltage 6 V)
Current consumption Approx. 90 mA with factory default settings
Power consumption Approx. 7 VA on input side (220 V AC side)
Dust and water proofing IP54 rating (for main unit)*2
Ambient conditions for operation -20 ºC to +50 ºC, 90 % RH or less (no condensation)
Dimensions and weight Approx. 175 mm (H) x 175 mm (W) x 40 mm (D) mm, approx. 780 g (incl. batteries)
SD card SD / SDHC (max. capacity 32 GB)*1
LED Two-color (red/blue) type for operation status indication
Supplied accessories Accelerometer PV-83D, Alkaline battery, IEC R6 (size AA) x 8,
Case x 1, 512 MB SD card x 1, Calibration Certificate
Accelerometer
Tri-axial
Accelerometer
PV-83D
(Cable: 1.5 m)
Rated sensitivity: 60 mV/(m/s2)
Frequency range: 0.5 Hz to 315 Hz
Usage temperature range: -20 ºC to +60 ºC (no condensation)
Waterproofing: IPX7
Dimensions and weight: Approx 67 mm (dia.) x 50.5 mm (D), approx. 450 g

*1 Use RION fully guaranteed products.
*2 Protection against harmful dust and water splashing from any direction.
Precautions regarding waterproofing
Before use, verify that the rubber side cover and the battery compartment lid are firmly closed.
To maintain the water and dust proof rating, internal packing replacement is required every two years (at cost).

Get Adobe' Reader'