ADCMT DC Voltage Current Source 6166 | NIHON DENKEI PHILIPPINES, Inc.
CONTENTS CONTENTS

ITEMLIST

Products Infomation

ADCMT DC Voltage Current Source 6166

6½-digit working standard suitable for calibration High-accuracy and highly stable with bipolar output

The 6166 is a DC voltage/current source that uses a PWM system in the reference voltage generation block for high accuracy, high stability and high resolution.
The DC voltage can be output over a wide range of ±10 nV to ±1199.999 V and the DC current over a wide range of ±1 nA to ±119.9999 mA. Especially with its high stability for DC voltage source of 25 ppm per 90 days (typical value), the 6166 can be widely used for calibration of high precision digital voltmeters or analog indicator instruments or as a generation source for a variety of tests.
In addition, the 6166 has a built-in function to generate thermal electromotive force of thermocouples according to the JIS table, enabling easy calibration of thermometers and other instrumentation systems.

DC Voltage Current Source

Model:

6166

Maker:

ADC CORPORATION

Features

  • Wide dynamic range and high resolution
    Voltage source: ±10nV to ±1200V
    Current source: ±1nA to ±120mA
  • High accuracy: ±35ppm/year (±25ppm/90days)
  • High stability: ±5ppm/24hrs
  • Smooth polarity switching by bipolar output
  • Memory up to 1000 data sets
  • JIS-compliant thermal electromotive force output function
  • Compliance voltage in 1mA/10mA range can be changed
    from ±120V to ±1200V. (factory option)

Specifications

Number of digits for generation
Output method Bipolar
Maximum output ±1200 V / ±12 mA
Voltage source range ±10 nV to ±1200 V
Voltage source accuracy (typical value)  0.0035 %
Current source range ±1 nA to ±120 mA
Current source accuracy (typical value) 0.01%
Thermal electromotive force output function -220 ℃ to +1820 ℃ (resolution: 0.1 ℃)
Output noise (typical range)  3 mVp-p
Settling time 1 s or less
Interface GPIB, USB,  BCD (factory option)
Memory 1,000

Get Adobe' Reader'

To view the pdf files, the Adobe Reader from Adobe Systems is required.

Researched by Shared Research Inc.
about Nihon Denkei Co., Ltd.