HIOKI LEAK CURRENT HiTESTER ST5541

Leak Current Measurement an Essential Part of Electrical Safety (for electrical devices)

To ensure the safe use of electrical products, electrical safety tests such as insulation resistance, withstand voltage, ground resistance, and leakage current must always be conducted. Hioki Leakage Current Testers comply with laws and standards regarding medical electrical equipment as well as non-medical electrical equipment, and can be used to measure leakage current in all types of electrical products from computers to medical equipment.

The ST5541 provides support for standard-compliant networks (excluding medical-use electrical devices).

LEAK CURRENT HiTESTER

Model:

ST5541

Maker:

HIOKI

Download Materials

For Standard- and Regulation-compliance Measurement of General-use Electrical Devices

 

There are various standards in place concerning networks (body simulated resistance), and a standard-compliant network is required in order to make measurements.
ST5541 provides standard support for standard-compliant networks (excluding medical-use electrical devices).

 

3157 Safety Standard for Hospital Electrical Equipment JIS T 1022:2006 Measurement

 

Combination of Instruments for Leak Current Testing and Safety Conductor Testing

The following are key parts of any safety inspection of electrical equipment:
•Leak current test: Measure with the ST5540 and ST5541.
•Safety conductor test (also known as a ground line resistance test or ground conductor test): Measure with the 3157.

 

The 3157 can also be used for conducting measurements under the JIS T 1022:2006 safety standard for hospital electrical equipment.

Key Feature

  • Compliance with Electrical Appliances and Materials Safety Act, JIS/IEC/UL standards
  • Uninterrupted polarity switching function dramatically reduces cycle time
  • Support for rated currents up to 20 A gives the instrument more than adequate capability for testing products designed to comply with new standards
  • Touch panel features simple, interactive operation
  • Communications functionality and external I/O support allow automatic testing on production lines

Specifications

Measurement

methods

Measurement of voltage drop across body simulated resistance points, Calculation and display of current values, True rms measurement, Measurement unit floats relative to instrument ground.
Measurement

modes

Leak current measurement, voltage measurement, safety conductor current measurement
Standards

compliance
(NW: Body

simulated resistance)

[NW-A] • Electrical Appliances and Materials Safety Act
[NW-C]
• Measurement of touch current and protective conductor current: IEC 60990:2016
• Electrical equipment for measurement, control, and laboratory use: IEC 61010-1:2010+ A1:2016
• Information technology equipment: IEC60950-1:2005+ A1:2009+ A2:2013
• Audio, video and similar electronic apparatus: IEC 60065:2014
• Personnel Protection Systems for EV: UL 2231-1:2012 (Amended 2016), UL-2231-2:2012 (Amended 2016)
[NW-D] • For UL: UL 1492:1996 (Amended 2013)
[NW-G] • Electrical equipment for measurement, control, and laboratory use; current measurement circuits in damp conditions: IEC 61010-1:2010+ A1:2016
Leak current

measurement

Ground leak current, 3 types of contact current, free current measurement, 3 types of enclosure leak current
Measurement

current

DC, AC (true rms, 15 Hz to 1 MHz), AC+DC (true rms, 15 Hz to 1 MHz), AC peak (15 Hz to 1 MHz)
Measurement

ranges

DC / AC / AC+DC mode: 50.00 mA/ 5.000 mA/ 500.0 μA/ 50.00 μA
AC peak mode: 75.0 mA/ 10.00 mA/ 1.000 mA/ 500.0 μA
Measurement accuracy
(current measurement)
DC measurement: ±2.0% rdg. ±6 dgt. (typ.)
AC / AC+DC measurement: ±2.0% rdg. ±6 dgt. (15 Hz to 100 kHz, typ.)
AC peak measurement: ±2.0% rdg. ±6 dgt. (15 Hz to 10 kHz, typ.)
Interfaces External I/O, USB 1.1 (communications), RS-232C
Functionality Automatic test, data storage for 100 target devices, clock, data backup, printed output (optional), etc.
Power supply 100/120/220/240 V AC (specify at time of order), 50/60 Hz, 30 VA rated power
Target device

power supply input

100 to 250 V AC, 50/60 Hz, Rated current input from terminal block: 20 A
Target device

power supply output

Output from terminal block: 20 A, Output from outlet: 15 A
Dimensions and mass 320 mm (12.60 in) W × 110 mm (4.33 in) H × 253 mm (9.96 in) D, 4.5 kg (158.7 oz)
Included accessories Test lead L2200 (Red ×1, Black ×1) ×1 set, Enclosure probe 9195 ×1, Power cord ×3, Spare fuse for measurement line ×1, Instruction manual ×1, CD-ROM ×1

ST5540, ST5541 List of functions (Network)

Item ST5540 ST5541
Network A
(Electrical Appliances and Materials Safety Act)
Network B (Medical-use electrical devices)
Network C (IEC 60990)
Network D (UL)
Network E (General-purpose 1)
Network F (General-purpose 2)
Network G (IEC 61010-1)

List of functions (Major functions)

Item ST5540 ST5541
Power on polarity switching function
Rated current 20 A
Function for checking for blown fuses
Frequency band switching
110% voltage output terminal (T3 terminal)
S10, S12, S13, E terminal

List of functions (Testing leakage current mode)

Item ST5540 ST5541
Earth leakage current
Touch current
Patient auxiliary current
Patient leakage current
Total patient leakage current
Free current
Enclosure – Earth leakage current
Enclosure – Enclosure leakage current
Enclosure – Line leakage current
Patient leakage current I
Patient leakage current II
Patient leakage current III

 