This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.


ITEMLIST
Products Infomation

Record Temperature and Humidity Simultaneously with Dual-Channel Data Logger
Hioki compact 1 or 2-channel data loggers feature a 60,000 data set/channel memory to measure temperature, humidity, voltage or instrumentation signals. The LR5001 logs temperature and humidity over 2 channels in as fast as 1 second intervals.
HUMIDITY LOGGER
Model:
LR5001
Maker:
HIOKI
Specification
Measurement
items |
Temperature 1 ch and Humidity 1 ch (Requires included or optional humidity sensor) |
Measurement
range |
Temperature: -40.0°C to 85.0°C, Humidity: 0 to 100% RH at sensor environment |
Basic accuracy | [Temperature] : ±0.5 °C (main unit + sensor accuracy, at 0.0 to 35.0 °C) [Humidity] : ±5 % rh (main unit + temperature / humidity sensor LR950x combination, at 20 to 30 °C / 10 to 50 % rh) Note: Basic accuracy is typical value, refer to the detailed catalog |
Storage capacity | Instantaneous value mode: 60,000 data/ch, Statistical value mode: 15,000 data/ch |
Recording interval | 1 to 30 sec., 1 to 60 min., 15 selections |
Recording modes | Instantaneous recording: at every recording interval Statistical value recording: Measure at one second intervals, and record the instantaneous, maximum, minimum, and average values within every recording interval |
Recording methods | One-time recording: Stop recording when the memory capacity is full Endless recording: Continue recording even when the memory capacity is full (old data is overwritten) Start: Logger button operation or scheduled time Stop: Logger button operation or scheduled time, or auto-stop when the memory capacity is full (at one-time recording) |
Other functions | Always backs up last recorded data; backs up recorded data and setting conditions when battery power is low Note: After batteries are replaced within 30 seconds, recording resumes automatically (Recording is interrupted during battery replacement) |
Waterproof and
dust-proof |
IP54 (EN60529) (with sensor connected, but not including sensor tip) |
Interfaces | Infrared optical communications with LR5091, LR5092-20 |
Power supply | LR6 (AA) Alkaline battery × 1, Battery life: Approx. 3 months (Instantaneous recording, with 1-minute interval and auto power saving, at 20 °C), Approx. 20 days (Instantaneous recording, with 1-second interval at 20 °C) |
Dimensions
and mass |
79 mm (3.11 in) W × 57 mm (2.246 in) H × 28 mm (1.10 in) D, 105 g (3.7 oz) |
Included
accessories |
LR6 (AA) Alkaline battery (built-in internal) × 1, Humidity Sensor LR9504 × 1, Instruction manual × 1, Operation guide × 1, Kickstand × 1 |