CONTENTS CONTENTS

ITEMLIST

Products Infomation

HIOKI CHEMICAL IMPEDANCE ANALYZER IM3590

For R&D applications of Electrochemical Components and Materials, Batteries, and EDLCs

Hioki LCR Meters and Impedance Analyzers range from 1 mHz to 3 GHz devices to suit a wide range of applications in the testing of electronic components. The IM3590 impedance analyzer is an advanced component tester especially suited to electrochemical impedance measurements and includes a variety of graphing functions that are needed for battery and chemical material testing.

CHEMICAL IMPEDANCE ANALYZER

Model:

IM3590

Maker:

HIOKI

Download Materials

Key Features

  • Broad 1 mHz to 200 kHz signal source range supports measurements of ion behavior and solution resistance
  • Continuous measuring and high-speed testing of LCR and sweep measurements with a single instrument
  • Measure internal impedance of batteries with no load
  • Perform high-speed sweep measurements in as little as 2 ms
  • Basic accuracy of ±0.05% is ideal for applications from component testing to R&D
  • Measure LCR impedance for Cole-Cole plots and equivalent-circuit analyses of electro-chemical components and materials

Battery internal resistance measurement and battery measurement function: Hioki IM3590 Chemical Impedance Analyzer

Specification

Measurement modes LCR mode, Continuous measurement mode (LCR mode / Analyzer mode), Analyzer mode (Sweeps with measurement frequency and measurement level, temperature characteristics, equivalent circuit analysis)
Measurement parameters Z, Y, θ, Rs (ESR), Rp, Rdc (DC resistance), X, G, B, Cs, Cp, Ls, Lp,
D (tanδ), Q, T, σ (conductivity), ε (dielectric constant)
Measurement range 100 mΩ to 100 MΩ, 10 ranges (All parameters are determined according to Z)
Display range Z, Y, Rs, Rp, Rdc, X, G, B, Ls, Lp, Cs, Cp, σ, ε :
±(0.00000 [unit] to 9.99999G [unit], Absolute value display for Z and Y only
θ : ±(0.000° to 180.000°), D : ±(0.00000 to 9.99999)
Q : ±(0.00 to 99999.9), Δ % : ±(0.0000% to 999.999%)
T : -10.0°C to 99.9°C
σ, ε :±(0.00000f [unit] to 999.999G [unit]
Basic accuracy Z: ±0.05% rdg. θ: ±0.03°
Measurement frequency 1 mHz to 200 kHz (5 digits setting resolution, minimum resolution 1 mHz)
Measurement signal level Normal mode:
V mode/CV mode: 5 mV to 5 Vrms, 1 mVrms steps
CC mode: 10 μA to 50 mArms, 10 μArms steps
Low impedance high accuracy mode:
V mode/CV mode: 5 mV to 2.5 Vrms, 1 mVrms steps
CC mode:10 μA to 100 mArms, 10 μArms steps
Output impedance Normal mode: 100 Ω, Low impedance high accuracy mode: 25 Ω
Display 5.7-inch color TFT, display can be set to ON/OFF
Measurement time 2 ms (1 kHz, FAST, display OFF, representative value)
Functions DC bias measurement, DC resistance temperature compensation (converted reference temperature is displayed), Temperature measurement, Battery mesurement (Automatic DC biasing system), Comparator, BIN measurement (classification), Panel loading/saving, Memory function
Interfaces EXT I/O (Handler), USB communication (high-speed), USB memory
Optional: Choose 1 from RS-232C, GP-IB, or LAN
Power supply 100 to 240 V AC, 50/60 Hz, 50 VA max.
Dimensions and mass 330 mm (12.99 in) W × 119 mm (4.69 in) H × 168 mm (6.61 in) D, 3.1 kg (109.3 oz)
Included accessories Power cord ×1, Instruction manual ×1, CD-R (Communication instruction manual and sample software [Communications control, accuracy calculation, and screen capture functionality] ) ×1

 