CONTENTS CONTENTS

ITEMLIST

Products Infomation

GRAPHTEC IWATSU Digital Oscilloscopes DS-5600A Series

Extended basic functionality, making it easier to use

Bandwidth100~500MHz
2 GS/s 4/2 Channels Max
Up to 5M points

Digital Oscilloscopes

Model:

DS-5600A Series

Maker:

GRAPHTEC IWATSU Test Instruments Co.,Ltd.

Features

DS-5600A Series All models 50 Ohm input compatible

Supports a wide variety of probes.

AUX OUT output terminal is equipped as standard.

In addition to the effective trigger signal output, the Pass/Fail determination function can output when the result becomes a Pass or Fail.

High Resolution

When measured at a sampling speed lower than the highest sampling speed, the data captured in the highest sampling can be averaged to capture waveforms, reduce random noise and increase vertical axis resolution to up to 12 bits equivalent equally.

Expanded average count

The average setting count was expanded from 256 times to 65536 times (maximum). This can effectively reduce non-periodic random signals from repeated measured signals.

Auto setup disable toggle function

Prevents the panel settings from being updated, even if the AUTO SET button is pressed by mistake.

Long memory up to 5M points (when all channels are used: 2.5m points/CH)

You can capture long waveforms while maintaining fast sampling.

4 item simultaneous judgment of waveform parameters/waveform mask judgment function

Automatic measurement of waveform parameters logging function (list display)
The pass/fail judgment can be performed automatically depending on the value of the mask or waveform parameter.
By performing 4 items at the same time, it is possible to set strict conditions.

Automatic measurement of waveform parameters logging function (list display)

Up to four waveform parameters can be logged. Logged data can be saved to a USB memory device. By importing data into Excel, etc., you can easily display the histogram (variable distribution) and trend display (change over time) of waveform parameters.

Full cycle parameter measurement*

Automatically measures the waveform of the captured full cycle (time axis). A large amount of data can be captured in the long memory to determine the maximum and minimum deflection widths. You can also analyze the interval that is pinched by the cursor.

* The waveform parameter of the vertical axis (voltage axis) displays the measured value once per sweep.

Enhanced trigger capability

The trigger function is enhanced and waveforms can be triggered under optimal conditions, even in logic signals and serial data signals with increased complexity.
Complex setting items performed by pattern triggering can also be set smoothly thanks to the operability of the touch screen.

Waveform operation function

Two waveforms can be added, subtracted, multiplied, and channel waveform frequency analysis (FFT) can be performed.
The DS-5600A/5600 Series supports differentiation and integration operations.
The operation waveform can be saved as data and set as the source for automated measurement of waveform parameters.

  • Addition and subtraction: evaluation of differential signals
  • Multiplication: Evaluation of power waveform by voltage x current
  • FFT: Analysis in the frequency range such as periodic noise and vibration

Probe selection function

When you select a model of an IWATSU ELECTRIC probe, the attenuation ratio and coupling are set automatically. The model number, vertical range bandwidth, and input coupling are displayed.

CH-to-CH skew adjustment function

The skew can be adjusted up to 500ns.*
For example, the phase adjustment between the current probe and the voltage probe allows for more accurate power waveform operations.

* When the horizontal axis is set to skew priority (there is a limit to the number of waveforms in the replay function).

Clear sweep function

You can reset the measured data such as replay waveform, averaging, persistence, automatic measurement of waveform parameters, pass/fail judgment, and logging.

High-speed remote waveform transfer

It achieves a remote transfer rate of about 100 times that of previous models.*
When collecting measurement inspection data remotely, you can reduce tact time and increase production efficiency.

* When the waveform display mode is turned OFF when the LAN interface is used. It is a measurement result under specific conditions, and the value is a reference. Actual transfer time will vary depending on the specifications of your PC.

Touchscreen independent keys/knobs

The 7.5-inch color LCD is equipped with a touchscreen function for intuitive operation.Although it is a compact size, the CH/MATH independent operation keys and knobs have been adopted to achieve one-action comfort.

Peak detection function

It can reliably capture noise (minimum 1ns pulse width) in very slow signals.

Rescale function

By measuring with current probes, shunt resistance, various sensors, etc., the obtained output voltage signal can be converted into each value and read directly.

Automatic measurement of waveform parameters (full cycle measurement of horizontal axis parameters)

You can choose from up to four of 26 waveform parameters to display. You can also display the maximum, minimum, and number of measurements. The cursor function also supports automatic measurement of a limited range. You can further expand the range of analysis by using the judgment function and logging function.

Simple Moving Average

Simple Moving Average (SMA) is a digital filter that can be configured per-channel to reduce noise by smoothing the sampling points at a specified width.

* It can also be used with single-shot signals that are not repeating signals.

The PNG format of the image data storage format is compatible with transparent type

Supports the transparent attributes of PNG, when creating a document using a PC, the waveforms can be stacked with transparency.

Specifications

Digital Oscilloscope “ViewGo II” DS-5600A Series Specifications

500MHzModel 350MHzModel 200MHzModel 100MHzModel
DS-5654A DS-5652A DS-5634A DS-5632A DS-5624A DS-5622A DS-5614A DS-5612A
Rise time
(Representative values)
750ps 1ns 1.75ns 3.5ns
Number of input channels 4 2 4 2 4 2 8 2
Frequency bandwidth (-3dB) 500MHz 350MHz 200MHz 100MHz
Maximum sampling speed 2GS/s (when channels are coupled), 1GS/s (when all channels are used)
Equivalent sampling speed 100GS/s
Peak detection resolution 1ns
Averaging function 2 to 65536 times (power of 2 steps)
Averaging
Count display function
Display the number of executions on the screen.
Memory length 5M points/ch (when 2 channels are coupled), 2.5M points/ch (all channels)
Vertical resolution 8 bits (when high resolution arithmetic is enabled: up to 12 bits)
Vertical resolution display function Display the number of resolution bits
Input voltage range 2mV/div ~ 10V/div(1MΩ)
2mV/div ~ 2V/div(50Ω)
Offset voltage 2mV/div ~ 50mV/div : ±1V、50.2mV/div ~ 500mV/div : ±10V、502mV/div ~ 10V/div : ±100V
DC gain accuracy ±(1.5% +0.5% full-scale)
Maximum input voltage ±400 Vpeak (1MΩ)、5 Vrms(50Ω)
Bandwidth limit filter Analog type:100MHz, 20MHz, 2MHz, 200kHz Analog type:20MHz, 2MHz, 200kHz
Digital type: Choose either LPF, HPF, or SMA, channel independent
Input coupling GND、DC 1MΩ、AC1MΩ、DC50Ω
Input impedance 1MΩ±1% //16pF、 50Ω±1%
Probe sense Auto search:1:1, 10:1, 100:1, 1000:1
Manual setting:1:1, 5:1, 10:1, 20:1, 50:1, 100:1, 200:1, 500:1, 1000:1, 2000:1
Time axis range 500ps/div ~ 50s/div 1ns/div ~ 50s/div 2ns/div ~ 50s/div 5ns/div ~ 50s/div
Standard probe SS-101R (with standard number of channels) SS-0130R (with standard number of channels)
Roll mode 50ms/div ~ 50s/div(100kS/s max)
Clock accuracy ±10 ppm
Power supply/power consumption AC100V ~ 240V(50Hz / 60Hz), AC100V ~ 120V(400Hz) / 95VA max(60W max)
Dimensions/Body weight Approx. 330Wx190Hx124D mm/Approx. 3.7kg
Performance warranty temperature 10 ~ 35℃
Operating temperature/
humidity/altitude
Temperature 0 to 40°C/Humidity 5 to 80% RH ≤ 30°C (no condensation),
55% RH or less 40°C (no condensation)/Altitude 2000m or less
Standard Accessories Probes (one for each channel), power cord (1), front panel cover (1), manual (CD-ROM) (1), user’s guide (1)

DS-5600A Series Trigger function

Contents
Trigger function Edge, Edge Alternate, Edge OR, Number of Pulses, Pulse Width, Period, Missing, TV, Pattern (OR, NOR, AND, NAND),
Serial (UART, SPI, I2C)
TV trigger standard NTSC、PAL、Custom
Line setting range selection up to 3000
Field selection 1、2、4、8
Pulse number trigger setting range 1 to 9999 events
Pulse width trigger time setting range 15ns~50s
Cycle trigger time setting range 40ns~50s
Missing (dropout)
Trigger time setting range
50ns~50s
Pattern trigger  OR、NOR、AND、NAND
Trigger source All channels
State HIGH、 LOW、 Don’t Care
Threshold level Independent setting for all channels
Serial trigger UART Trigger selection START, STOP,Parity error, data pattern
Comparison data length 5 to 8 bits
Bit rate 1000bps~1Mbps(Set in 100bps units)
Signal source CH1~CH4, EXT(2-channel device is CH1, CH2, EXT)
Serial trigger UART Trigger selection Data pattern
Comparison data length 4 to 64 bits
CS selection Positive logic/Negative logic,Idle time specification when there is no CS
Signal source CH1~CH4, EXT(2-channel device is CH1, CH2, EXT)
* CH1 input is for SCK signal input only: up to 20MHz
Serial trigger I2C Trigger selection START, STOP, RESTART, NACK,Data pattern
Comparison data length 7-bit/10-bit address, 1 to 5 bytes,EEPROM read, byte (shift comparison available)
Address mode Choose from 7-bit/10-bit/EEPROM read
Signal source CH1~CH4, EXT(2-channel device is CH1, CH2, EXT)
Trigger source All channels, EXT (±0.5V), EXT10 (±5.0V), line
Trigger slope Positive, negative
Coupling AC, DC, high frequency removal, low frequency removal, noise removal

Display/Resolution

Contents
Display 7.5-inch color TFT LCD (touch screen)
Resolution VGA:640 × 480 pixels
Display format YT, XY, XY Trigger
Vector connection Sample point interpolation display, dot display
Analog persistence Solid tone display, spectrum display
Persistence Display time 100ms, 200ms, 500ms, 1s, 2s, 5s, 10s,Infinite
Save internal waveform(REF memory) 5 waveforms
Front panel settings save Five settings can be saved in the internal memory or USB memory device
AUTO SETUP function Can be set to Enabled or Disabled

Parameter measurement, cursor, zoom, math, replay

Contents
Parameter measurement Maximum value, minimum value, peak value, RMS value, cycle RMS value, average value, cycle average value, top, base, top-base, rising overshoot, falling overshoot, rising time 20-80%, falling time 80-20%, rising time 10-90%, falling time 90-10%, frequency, cycle, number of pulses (positive), number of pulses (negative), pulse width (positive), pulse width (negative), duty ratio, integral, skew (positive/negative), skew @ level
Number of simultaneous measurements Maximum 4 parameters
Statistical values display maximum value, minimum value, number of measurements
Logging items Time, parameter measurement results (conditions A, B, C, and D), Pass/Fail judgment results
Output destination When recording: Pop-up screen, internal memory (up to 86,400)
after recording: USB memory device
Pass/Fail Judgment mode: parameter judgment or mask determination, judgment result: USB save, BEEP sound,
pulse output, logging
Page search function: select Pass or Fail, search forward or backward
Cursor Time, Amplitude, Time & Amplitude, Value at Cursor
Zoom Press the Zoom button on the front panel to display the magnified waveform in a different grid
Math function Addition, subtraction, multiplication, differentiation, integration, FFT (up to 8k points, rectangular, Hanning, flat top window function)
Double operation by either differential, integral, or FFT from the result of either addition, subtraction, or multiplication (9 patterns)
Rescale a * x + b (x: input voltage, a, b: user-defined)
unit conversion volt, ampere, watt, °C, no display
Replay Automatic waveform recording, up to 2048 waveforms can be saved and replayed
Frequency counter 6-digit
Interface USB 2.0HS compatible (device, host), LAN (100Base-TX),GPIB (factory option DS-576), AUX interface (connector for external option)
AUX OUT Trigger output and pass/fail judgment output can be switched
Waveform data save Save to USB memory device with binary, ASCII, Mathcad, arithmetic (ASCII), and arithmetic (Mathcad)
Hard copy output Save the image to a USB memory device in TIFF, BMP, or PNG (Transparent) format
or output to a PictBridge® -compatible printer
Calibration signal output Square wave 1kHz, 3Vp-p

Option

Contents
AUX IO CH1/CH2 output DS-577
(Factory option)
AUX IO1: Output CH1 input signal with offset voltage added
AUX IO2: Output CH2 input signal with offset voltage added
AUX IO CH1 output DS-578
(Factory option)
Aux IO1: Output CH1 input signal with offset voltage added
GPIB interface DS-576
(Factory option)
GPIB: IEEE488.2
Probe power supply DS-579 IWATSU ELECTRIC active probe power supply 2 Series