CONTENTS CONTENTS

ITEMLIST

Products Infomation

SOKEN Partial Discharge Tester DAC-PD-7

Magnet wires, printed circuit boards, connectors, semiconductors, small motors, electrostatic chucks and insulating materials

Since DAC-PD-7 contains the necessary circuits for PD testing, such as the test power supply, detector, coupling capacitor and calibrator, it can easily perform partial discharge tests with only the main unit. Partial discharge testing to judge the quality of magnet wires, printed circuit boards, connectors, semiconductors, small motors, electrostatic chucks and insulating materials can be easily performed. PD tests according to the international standard (IEC 60270, 60664-1. 61730-2) are available with the PC software.

Partial Discharge Tester

Model:

DAC-PD-7

Maker:

SOKEN ELECTRIC CO.,LTD.

Download Materials

PC Software

Voltage-Discharge Quantity Characteristic Test

AC test voltage is linearly increased to the max test voltage/PD Inception voltage, and automatically decreased.
PD inception and extinction voltages are determined according to a specified Threshold value (pC).

 

TIME-Discharge Quantity Characteristic Test

In this mode, two levels of voltage “V1” and “V2” are applied continuously.
“V1” is held for “T1” seconds and “V2” is held for “T2” seconds.
PD inception and extinction voltages are determined according to a specified Threshold value (pC).
If no discharge exceeding the criterion value occurs during the last 5 seconds of the holding time “T2”, it is judged as “PASS”, and if it occurs, it is judged as “FAIL”.

Features

All-IN-ONE UNIT
The tester contains 3000V (or 5000V) testing power supply, coupling capacitor, detector and calibrator.
EASY TO USE
No need cumbersome wiring.
AUTO CALIBRATION
Calibration can be performed by simply pressing AUTO CAL key.
AUTO ADJUSTMENT RANGE
The measurement range is automatically adjusted according to DUT.
EASY-TO-UNDERSTAND DISPLAY OF MEASUREMENT STATUS
The measurement can be conducted while observing discharge waveform.
The applied voltage profile can be observed simultaneously.
AUTO MEASUREMENT WITH SOFTWARE
RS232C interface is provided. Automatic measurement with the software is possible.

Specifications

Input Voltage AC100V/115V/220V/240V ±10%, 50/60Hz
Output Voltage AC 0 – 5000V rms
Voltage Ramp Rate 25V/s, 50V/s
(10V/s, 25V/s, 50V/s and 100V/s with software)
PD Resolution 0.01pC
(Partial discharge in device <1pC at AC3000V)
Frequency 50Hz, 60Hz
Waveform Sine wave
(not depend on waveform of Input Power)
Distortion of Waveform <3%
Fluctuation of Voltage <1%
Maximum Load Current 10mA
Maximum Capacitance Load 5000pF
Gain 0~80dB
Range 1000pC, 10000pC, 100000pC
Response Occurrence Frequency 10 – 9000PPS (Rate)
Calibrator Charge 100pC, 1000pC
Calibration Pulse Injection Capacitor 50pF
Calibration Pulse Voltage 2V, 20V
Calibration Pulse Generation Frequency 50PPS
Internal Coupling Capacitor 1000pF
Interface RS232C
Size 430 × 380 × 200 (W × D × H)
Weight 15 kg
Standard Accessory Measurement cable, Power cord, PC software,
serial cable
Option Shield Case, Electrode, USB-RS232C converter

Get Adobe' Reader'

To view the pdf files, the Adobe Reader from Adobe Systems is required.

Researched by Shared Research Inc.
about Nihon Denkei Co., Ltd.