SIGMA Vibration Monitor VM Series | PT. NIHON DENKEI INDONESIA Jakarta
CONTENTS CONTENTS

ITEMLIST

Products Infomation

SIGMA Vibration Monitor VM Series

Vibrometer and data recording combined with, Analysis Function added

•Vibration severity standard JIS B 0907 or equivalent
•Simultaneous measurement and recording of acceleration, velocity and displacement
•FFT (spectrum) and time waveform functions
(VM-2001H, VM-1001H standard equipment)
•USB communication with PC and hard copy of screen available
•Automatic saving of measurement data to microSD card
•Battery/AC adapter can be used together

Vibration Monitor VM Series

Maker:

SIGMA Electronics

Feature

Features of VM-1001H/S

•Compact and lightweight 3.5″ LCD type
•Simple type with 1-channel measurement
•Frequency analysis function (equipped on type-H)
•Vibration waveforms can be recorded using the time waveform function (equipped on type-H)
•Bearing diagnostics can be performed by setting bearing specifications (equipped on type-H)

 

Features of VM-2001H/S

•Large, easy-to-read 5.7″ LCD screen
•Simultaneous 2-channel measurement and diagnosis
•2-channel split and overlay display
•Frequency analysis function (equipped on type-H)
•Vibration waveforms can be recorded using the time waveform function (equipped on type-H)
•Bearing diagnostics can be performed by setting bearing specifications (equipped on type-H)

Specifications

Specification VM-2001H VM-2001S VM-1001H VM-1001S
Vibration

input channels

2ch 1ch
Signal

input

CCLD 24V 2mA, BNC connector
*A charge converter is required when using Charge-output type pickups.
Measuring

range

Acc. Peak Greatest value : 280m/s²
Acc. RMS Root mean squared value : 200m/s²
Acc. C.F. Acc. C.F. = Acc. Peak / Acc. RMS
Vel. RMS Root mean squared value : 400mm/s at 80Hz
Disp. RMS Root mean squared value : 800μm at 80Hz
Disp.

EQP-P

Equivalence p-p value : 2240μm at 80Hz
Frequency

range

Acceleration 10Hz to 10kHz ±1.5dB (at 80Hz)
10Hz to 20kHz -3.0dB (at 80Hz)
Velocity 10Hz to 1kHz (at 80Hz, based on JIS B 0907-1989)
Displacement 10Hz to 1kHz (at 80Hz)
Vibration value resolution Acc.P:1.3% of vibration measurement range
Filter 【HPF】 3Hz / 10Hz / 1kHz Cutoff characteristics:-24dB/oct
【LPF】 200Hz / 1kHz / 5kHz / 20kHz Cutoff characteristics:-24dB/oct
Analytical

function

OverAll

Vibration

analysis

Meas. items : Acc. Peak, Acc. RMS, Acc. C.F. Vel. RMS, Disp. EQP-P
Simultaneous display
Disp. EQP-P can be switched with Disp. RMS
Vibration history: previous vibration data (up to 400 points) can be displayed without an external device
FFT

analysis

Meas. items : Acc. RMS, Vel. RMS, Disp. EQP-P, Envelope
Max. analysis frequency : 200, 500, 1k, 2k, 5k, 10k, 20k (Hz)
Number of sampling data : 64, 128, 256, 512, 1024, 2048, 4096, 8192
*Combination of two items from the above is indicated as “Frequency resolution”.
Timewave

form

Meas. items : Acc.
Max. analysis frequency : 200, 500, 1k, 2k, 5k, 10k, 20k (Hz)
Number of line : 25, 50, 100, 200, 400, 800
*Combination of two items from the above is indicated as “Frame time axis length”.
Capable of FFT calculation of acquired waveforms
Display 5.7″ color LCD
When measuring 2 channels, “Split display,” “Overlapping display,” or “Switching display” can be selected.
3.5″ color LCD
Display

lamp

Charging / Charging complete
Panel

interface

LCD touch panel
Switching gain

function

Automatic/manual selectable
Data log

function

Equipped in all analysis modes, arbitrary data storage intervals can be set.
USB port USB mini-B
micro SD card

slot

Data log (.csv), Hard copies of screens are available
Environment Temperature : 10 to 40℃ (Non-condensing)
Humidity : 20 to 80%RH
*Only body
Power

supply

Li-ion battery (3 hours continuous operation)
AC adapter (AC100 to 240V ±10% 50/60Hz)
Dimension (mm) 187.5(W) × 53.5(D) × 130(H) 180(W) × 45(D) × 100(H)
Weight 800g 350g
Standard

accessories

1. AC adapter (1 pcs)
*Choose one of the following
•Japanese specification: UN318-0628
(AC100 to 240V±10%, 50 to 60Hz)
•Overseas specifications: UI318-06
(AC100 to 240V±10%, 50 to 60Hz)
Plug shape A type for both models.
AC cord conversion method for overseas specifications.
The AC cord power cable must be provided by the customer.

2. CD-ROM (1 pcs)
Manual (PDF), Installation Software

1. AC adapter (1 pcs)
*Choose one of the following
•Japanese specification: UN312-5920
(AC100 to 240V±10%, 50 to 60Hz)
•Overseas specifications: UI318-06
(AC100 to 240V±10%, 50 to 60Hz)
Plug shape A type for both models.
AC cord conversion method for overseas specifications.
The AC cord power cable must be provided by the customer.

2. Stylus (1 pcs)
Type : PDA-PEN16
Manufacturer : Sanwa Supply

3. CD-ROM (1 pcs)
Manual (PDF), Installation Software

4. Vibration Sensor (1 pcs)
Type : SAF12C
Sensitivity : 10mV/(m/s2)
Cable length : 2.5m (curled cord)   Isolation : Support
Waterproof : Support

5. Sensor fixing magnet (1 pcs)
Type : SHXU16
Adsorption force : 58.8N

Optional 1. Vibration Sensor
Type : SAF12C
Sensitivity : 10mV/(m/s2)
Cable length : 2.5m (curled cord)   Isolation : Support
Waterproof : Support

2. Sensor fixing magnet
Type : SHXU16
Adsorption force : 58.8N

Get Adobe' Reader'