KIKUSUI AC/DC Hipot Tester TOS5401 & TOS5402 [NEW]

TOS5400 series electrical safety testers are designed to perform two of the four safety tests required for electrical products

Withstanding voltage and insulation resistance tests. With an output of 5 kV / 100 mA (AC) and 6 kV / 10 mA (DC), these devices can perform withstanding voltage and insulation resistance testing of electronic equipment and components in accordance with safety standards such as IEC, EN, BS, VDE, UL, CSA, GB, and JIS, as well as the requirements of the Japanese Electrical Appliance and Material Safety Act. TOS5400 series is also equipped with a highefficiency PWM switching amplifier that maintains stable output voltage even if there are AC line voltage or frequency fluctuations, enabling reliable evaluation even in regions with unstable power supply.

AC/DC Hipot Tester

Model:

TOS5401 & TOS5402

Maker:

KIKUSUI

Download Materials

Feature Details

  • PWM amplifier provides highly stable output
  • AC withstanding voltage test: 5 kV / 100 mA (500 VA) (TOS5401/TOS5402)
  • DC withstanding voltage test: 6 kV / 10 mA (50 W) (TOS5401/5403)
  • Touch panel display
  • LAN, USB, and RS-232C as standard
  • Shorter takt time thanks to faster output control and measurement

Specification

AC output section (ACW only)

Item TOS5401 TOS5402
Output range 0.05kV to 5.00kV
Output setting accuracy ±(2% of setting + 20V) (at no load)
Output setting range 0.000kV to 5.500kV
0.000kV to 1.500kV (When Time Unit is set to Cycle and Phase Shift is set to 0°)
Output resolution 1V
Max. rated output *1 500VA (5kV/100mA)
Max. rated voltage 5kV
Max. rated current 100mA (when the output voltage is 0.5kV or higher)
Transformer rating 500VA
Output voltage waveform*2 Waveforms Sine
Phase shift The output voltage waveform can start with a phase shifted approx. 90° from the zero-crossing point.
Distortion 3% or less. (when the output voltage is 0.5kV or higher and no load or a pure resistive load is connected)
Frequency 50Hz/60Hz
Frequency setting accuracy ±0.5% (excluding during voltage rise time)
Voltage regulation ±10% or less (when changing from maximum rated load to no load)
Input voltage variation ±0.3%(5kV at no load; power supply voltage: 90V to 250V)
Short-circuit current 200 mA or more (output voltage 1.0 kV or higher)
Output method PWM switching

 

Common specifications

Item TOS5401 TOS5402 TOS5403
Start voltage function The initial voltage at test start can be set to 1% to 99% of the test voltage. Resolution: 1%
Disabled if the Time Unit is set to Cycle. n/a
Limit voltage function The upper limit of the test voltage can be set.
ACW: 0.000kV to 5.500kV, DCW: 0.000kV to 6.200kV
ACW: 0.000kV to 1.500kV (When Time Unit is set to Cycle and Phase Shift is set to 0°)
Output voltage monitor function If the output voltage deviates more than ±(350V) form the set value, the output is turned off, and the protection function is activated.

Digital voltmeter

Item TOS5401 TOS5402 TOS5403
Measurement range AC/DC: 0.000kV to 6.500kV AC: 0.000kV to 6.500kV DC: 0.000kV to 6.500kV
Display □. □□□ kV
Accuracy V < 500V: ±(1.5% of reading + 10V)
V ≥ 500V: ± 1.5% of reading
Response*1 ACW: True rms/ Mean-value response rms display can be switched.
DCW and IR: Mean-value
Mean-value
Hold feature Holds the measured voltage value at the end of the test while the PASS/FAIL judgment result is displayed.

 