Electrical specifications
|
| Rated power |
3750W |
2800W |
| Rated current |
28 Arms / 84 peaks |
18.75 Arms / 56.25 Apeak |
| Rated voltage |
50-480Vrms / 700Vdc |
| Rated frequency |
DC, 40~440Hz (CC CP Mode) / DC~440Hz (LIN, CR, CV Mode) |
| OPP ( Over Power Protection ) |
3937.5Wrms or arbitrary set value |
2940Wrms or arbitrary setting |
| OCP ( over current protection ) |
29.4Arms or arbitrary set value |
19.687Arms or arbitrary set value |
| OVP ( overvoltage protection ) |
367.5Vrms / 525Vdc |
| OTP ( Over Temperature Protection ) |
can be |
CC mode
and
linear CC mode |
range |
0 to 28A |
0 to 18.75A |
| resolution |
0.5mA |
0.3125mA |
| Accuracy *2) |
±(0.2% of setting + 0.4% of range) @50/60Hz
±(1% of setting + 1% of range) @DC
±(1% of setting + 2% of range) @400Hz |
| CR mode |
range |
2.0 to 40kΩ |
3.2 to 64kΩ |
| Resolution *1) |
0.0083333mS |
0.0052083mS |
| Accuracy |
±0.4% of (setting + range) @50/60Hz, ± (1% of setting + 4% of range) |
| CV mode |
range |
50-350Vrms / 500Vdc |
| resolution |
0.1V |
| Accuracy |
±(0.2% of setting + 0.2% of range) |
| CP mode |
range |
3750W |
2800W |
| resolution |
0.1W |
| Accuracy *2) |
±(0.2% of setting + 0.2% of range) @50/60Hz/DC
±(0.2% of setting + 4% of range) @400Hz |
CF
(CC mode, CP mode) |
range |
√2 to 5 |
| resolution |
0.1 |
| Accuracy |
(1% / lrms) + 2% FS . |
PF
(CC mode, CP mode) |
range |
0-1 (lag or lead) |
| resolution |
0.01 |
| Accuracy |
2% FS . |
| UPS efficiency measurement |
Non-linear mode |
| operating frequency |
Auto ; 40-440Hz |
| Current range |
0 to 28A |
0 to 18.75A |
| PF range |
0 to 1 |
PV system efficiency measurement
(inverter: 80% distortion) |
Resistance value + non-linear mode |
| operating frequency |
Auto ; 40-440Hz |
| Current range |
0 to 28A |
0 to 18.75A |
| resistance range |
2.0 to 40kΩ |
3.2 to 64kΩ |
UPS backup time measurement
(CC mode, LIN mode, CR mode, CP mode) |
UVP (VTH) |
50-350Vrms / 500Vdc |
| UPS Back-Up Time |
1 to 99999 sec. (>27H) |
Battery discharge function measurement
(CC mode, LIN mode, CR mode, CP mode) |
UVP (VTH) |
50-350Vrms / 500Vdc |
| battery discharge time |
1 to 99999 sec. (>27H) |
| UPS offline transfer time measurement |
Current range |
0 to 28A |
0 to 18.75A |
| UVP (VTH) |
2.5V |
| Time range |
0.15mS to 999.99msec. |
| turbo mode |
Maximum current (ON) |
56.0 Arms |
37.5 Arms |
| Maximum current (OFF) |
28.0 Arms |
18.75 Arms |
| Fuse test mode |
Trip time & non-trip time |
(Turbo ON) 0.1 to 1.0 sec. |
| (Turbo OFF) 0.1 to 9999.9 sec. |
| Measurement accuracy |
±0.006sec. |
| number of repetitions |
0 to 255 |
| Short/ OPP /OCP test function |
short circuit time |
(Turbo ON) 0.1S~1sec./ (Turbo OFF) 0.1S~10sec. Or Cont. |
| OPP /OCP step time |
(Turbo ON) 100msec., up to 10 Steps / (Turbo OFF) 100msec. |
| OCP Istop |
(Turbo OFF) 28.0 Arms
(Turbo ON) 56.0 Arms |
(Turbo OFF) 18.75 Arms
(Turbo ON) 37.5 Arms |
| OPP Pstop |
(Turbo OFF) 3750W
(Turbo ON) 7500W |
(Turbo OFF) 2800W
(Turbo ON) 5600W |
Programmable
inrush current simulation *4) |
starting current |
0 to 56A |
0 to 37.5A |
| step time |
0.1mS to 100mS |
| stop current |
0 to 28A |
0 to 18.75A |
Programmable
surge current simulation *4) |
S1&S2 current |
0 to 56A |
0 to 37.5A |
| T1&T2 time |
0.01S~0.5S |
| S3 current |
0 to 28A |
0 to 18.75A |
| T3 time |
0.01S to 9.99Sec. or Cont. |
| voltage display |
range |
500V |
| resolution |
0.01V |
| Accuracy |
±0.1% of (reading + range) |
| parameter |
Vrms, VMax/Min, +/-Vpk |
| current display |
range |
14 Arms / 28 Arms |
9.375 Arms / 18.75 Arms |
| resolution |
0.3mA/0.6mA |
0.2mA/0.4mA |
| Accuracy *2) |
±(0.1% of reading + 0.1% of range) @50/60Hz
±(0.4% of reading + 0.4% of range) @DC
±(0.4% of reading +4% of range) @400Hz |
| parameter |
Irms, VMAX/MIN, +/-Ipk |
| power display |
range |
3750W |
2800W |
| resolution |
0.0625W |
0.05W |
| Accuracy *2) |
±(0.2% of reading + 0.2% of range) @50/60Hz/DC
±(0.2% of reading + 4% of range) @400Hz |
| VA display |
Calculated from current and voltage measurement results |
| Power factor display |
range |
+/- 0.000~1.000 |
| Accuracy |
±(0.002±(0.001/PF)*F) |
| frequency display |
range |
DC, 40~440Hz |
| Accuracy |
0.002 |
| Other parameter display |
VA, VAR, CF_I, Ipeak, Imax, Imin, Vmax, Vmin, IHD, VHD, ITHD, VTHD |
| power on loading |
With power-on loading function |
| Load ON / OFF angle specification |
Can be specified between 0 and 359 degrees for the angle at which the load starts or stops |
| Master-slave function (3-phase application) *3) |
Up to 7 slave units for 1 master unit |
| External analog input (optional) |
FS / 10Vdc, resolution 0.1V |
| External sync input |
TTLs |
| Voltage monitor output (isolated) |
±500V / ±10V |
| Current monitor output (isolated) |
±84Apk / ±10Vpk |
±56.25Apk / ±10Vpk |
| Option interface |
GPIB, RS-232, LAN, USB |
|
General specifications
|
| Input power rated voltage |
AC 100V/200V ±10%, 50/60Hz |
| operating temperature |
0 to 40°C |
| operating humidity |
20 to 85% RH (non-condensing) |
| Storage temperature |
-20 to 70°C |
| storage humidity |
20 to 85% RH (non-condensing) |
| Dielectric strength (between input and FG) |
AC1500V, 1 minute |
| Dielectric strength (between input and load terminals) |
AC3000V, 1 minute |
| Dimensions (W×H×D) |
440x188x558mm |
| mass |
33.5kg |
27.5kg |