This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.


ITEMLIST
Products Infomation

3-Phase 4-Wire Power Quality Analyzer Complies with IEC61000-4-30 Class S Standard and On-Screen Guidance to Provide Easy Setup and Testing
Hioki portable power meters and analyzers are best-in-class power measuring instruments for measuring single- to three-phase lines with a high degree of precision and accuracy. The PQ3100 is a comprehensive yet easy-to-use power quality analyzer for monitoring and recording power supply anomalies, allowing their causes to be quickly investigated, and also for assessing power supply problems such as voltage sags, flicker, harmonics, and other electrical problems.
POWER QUALITY ANALYZER
Model:
PQ3100
Maker:
HIOKI
How to Use the Hioki PQ3100 Power Quality Analyzer: A Simple Solution for Wiring and Configuration
The Hioki PQ3100 Power Quality Analyzer provides an intuitive QUICK SET navigation system to simplify wiring and configuration procedures for measuring 3-phase systems. Connection errors such as sensor clamp reversal or phase sequence are automatically detected along with suggestions on how to correct the settings.
How to Use the Hioki PQ3100 Power Quality Analyzer: Manage Power Quality with Greater Convenience with the PQ ONE Package App
The free PQ ONE application software for PQ3100 Power Quality Analyzer allows you to analyze captured data in fine detail and create regular power quality management reports easily and flexibly. Investigate the causes of power anomalies in the comfort of your office using your PC.
Remote Monitoring of Site Power Supply Quality [Gennect One SF4000]
Get measurement results from the job site in real-time, and display/save them on your PC. Immediately and automatically transfer power quality event measurement files to your PC. The power quality analyzer itself can be remotely controlled from your PC to change event conditions.
Value Kit Model: POWER QUALITY ANALYZER KIT PQ3100-91
Model Number (Order Code) PQ3100-91
Kit contents: CT7136 AC current sensor (600 A) ×2, PQ3100 main unit, 2GB SD memory card Z4001, Carrying case C1009
Note: Choose from Value Kits for additional savings.
Value Kit Model: POWER QUALITY ANALYZER KIT PQ3100-92
Model Number (Order Code) PQ3100-92
Kit contents: CT7136 AC current sensor (600 A) ×4, PQ3100 main unit, 2GB SD memory card Z4001, Carrying case C1009
Note: Choose from Value Kits for additional savings.
Value Kit Model: POWER QUALITY ANALYZER KIT PQ3100-94
Model Number (Order Code) PQ3100-94
Kit contents: AC flexible current sensor CT7045 (6000 A) ×4, PQ3100 main unit, 2GB SD memory card Z4001, Carrying case C1009
Note: Choose from Value Kits for additional savings.
Feature
- Record data including voltage, current, power, harmonics and flicker simultaneously along one time axis.
- Quick Setup: Easy-to-understand on-screen guide to measurement procedures
- Measure DC current accurately over long periods (with AC/DC automatic zero current sensor)
- Bundled PQ ONE application software makes it easy to create reports
- Supply power directly to the connected current sensor
- Capture all power anomalies, including outages, voltage sags, and frequency fluctuations, while recording trend data.
- Record waveforms up to 1 second before and 10 seconds after an anomaly occurs.
- Send measured values to a Hioki data logger using an adapter compatible with Bluetooth® wireless technology (LR8410 Link compatible products), Ver. 2.0 and later.
How to Use a Power Quality Analyzer: Recording Grid Power Quality
Spesification
Types
of Measurement |
Single-phase 2-wire, Single-phase 3-wire, Three-phase 3-wire or Three-phase 4-wire plus one additional input channel CH4 for voltage/current, (all AC/DC measurement channels) |
Voltage
Range |
Voltage measurement: 1000.0 V rms or DC, Temporary measurement 2,200 kV peak |
Current
Range |
50,000 mA AC to 5,0000 kA AC, 10,000 A DC to 2,0000 kA DC (depending on the current sensor used) |
Power
Range |
50,000 W to 6,0000 MW (automatically determined based on current usage range) |
Basic
Accuracy |
Voltage: ±0.2% of nominal voltage, Current: ±0.1% rdg. ±0.1% fs + current sensor accuracy, Active power: DC ±0.5% rdg. ±0.5% fs + current sensor accuracy, AC ±0.2% rdg. ±0.1% fs + current sensor accuracy |
Measurement
details |
1. Transient over voltage: 200 kHz sampling 2. Frequency cycle: Calculated as one cycle 3. Voltage (1/2) RMS, Current (1/2) RMS: one cycle calculation is updated every half cycle 4. Voltage swell, Voltage dips, Voltage interruption, RVC (Ver.2.0 or later): Voltage (1/2) RMS calculation 5. Inrush current: half cycle calculation: Calculated as the RMS value of the current for current waveform data sampled every half cycle. 6. Frequency 200 ms: Calculated as 10 or 12 cycles 7. Frequency 10 seconds: Calculated as the entire cycle time during the specified 10-second period 8. Voltage waveform peak, Current waveform peak 9. Voltage, Current, Active power, Apparent power, Reactive power, Active energy, Apparent energy, Reactive energy, Energy cost, Power factor, Displacement power factor, Voltage unbalance factor, Current unbalance factor 10. Voltage peak factor, Current peak factor 11. Harmonic/ Harmonic phase angle (voltage/current), Harmonic power: 0 th to 50 th order 12. Harmonic voltage-current phase angle: 1st to 50th order 13. Total harmonic distortion factor (voltage/current) 14. Inter harmonic (voltage/current): 0.5 to 49.5 order 15. K factor (multiplication factor) 16. IEC Flicker, Δ V10 Flicker |
Catatan | Maximum recording interval: 1 year, Maximum number of events that can be recorded: 9999 × 365 days |
Interfaces | SD/SDHC memory card, RS-232C (for communication/ LR8410 link), LAN (HTTP/ FTP server/ Send email), USB 2.0 (for communication) |
Logger
Connectivity |
Send measured values wirelessly to the logger using a Bluetooth® wireless technology serial conversion adapter. (Supported devices: Hioki LR8410 compatible loggers), Ver. 2.0 and later |
Appearance | LCD warna TFT 6,5 inci (640 × 480 titik) |
Power
Resources |
AC Adapter Z1002 (100 V to 240 V AC, 50/60 Hz, rated current 1.7 A), Battery Z1003 (Continuous use: 8 hours, Charging time: Max. 5 hours 30 m with AC adapter) |
Dimensi
dan massa |
300 mm (11.81 in)W × 211 mm (8.31 in)H × 68 mm (2.68 in)D, 2.5 kg (88.2 oz) (including battery) |
Accessories
included |
Instruction manual ×1, Measuring guide ×1, L1000-05 tension cable ×1 set (Red/Yellow/Blue/Gray/Black, Alligator clip ×5, Spiral tube ×5), Color clip (to identify the color of the sensor clamp) ×1 set, Spiral tube ×5, AC adaptor Z1002 ×1, Strap ×1, USB cable (length 1m 3.28ft) ×1, Battery Z1003 ×1, PQ ONE (software, CD) ×1 |