This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.


ITEMLIST
Products Infomation

0 to 0.2 kV/5 kV AC Insulation and Hipot Tester
Insulation / WIithstanding HiTESTER 3159
Model:
3159
Maker:
HIOKI
Fitur
-
Pengujian insulation berkelanjutan (500 V/1000 V) dan withstand voltage (kapasitas transformator 500 VA)
-
Insulation to withstand series test atau withstand to insulation series test dalam mode otomatis, atau uji individu dalam mode manual
-
Menyimpan hingga 10 pengaturan pengujian masing-masing untuk withstanding dan insulation test mode
-
External I/O, RS-232C interface, Status output (relay contacts)
Spesifikasi
Monitor function |
Output voltage, detected current, insulation resistance, Refresh rate: 2 times/s
|
Power supply | 220 V AC, (50/60 Hz), 800 VA max. |
Dimensions and mass
|
320 mm (12.60 in) W × 155 mm (6.10 in) H × 330 mm (12.99 in) D,
|
21.5 kg (758.4 oz) | |
Accessories |
H.V. Test Lead 9615 (high voltage side and return, 1 each) ×1, Power cord ×1, Instruction manual ×1, Spare fuse ×1
|
[Withstanding test] | |
Testing voltage |
0 to 2.5 kV / 0 to 5.0 kV AC, 2 range configuration 500 VA (30 minutes rated)
|
Voltage setting | Manual setting |
Waveform/Frequency |
Same as the power supply waveform, synchronized with the power supply
|
Current measurement |
0.01 mA to 120 mA, True RMS rectified (digital display)
|
Voltage meter |
Accuracy: ±1.5 % f.s. (digital), ±5 % f.s. (analog, f.s.=5 kV)
|
Current measurement |
0.01 mA to 120 mA, (Average value rectified, effective value digital display)
|
Measurement range |
2 mA/8 mA (0.01 mA resolution), 32 mA (0.1 mA resolution), 120 mA (1 mA resolution)
|
Voltage meter | Digital, Accuracy: ±1.5 % rdg. (f.s.=5.00 kV) |
Judgment function |
Window comparator method (Digital setting)
|
[Insulation test] | |
Testing voltage |
500 V DC, 1000 V DC, Unloaded voltage: 1 to 1.2 times rated voltage
|
Rated testing current |
1 to 1.2 mA, Shorted current: 4 to 5 mA (at 500 V), 2 to 3 mA (at 1000 V)
|
Measurement range/accuracy
|
0.5 MΩ to 999 MΩ (at 500 V), and 1 MΩ to 999 MΩ (at 1000 V): ±4 % rdg.,
|
1000 MΩ to 2000 MΩ: ±8 % rdg | |
Judgment function |
Window comparator method (Digital setting)
|
[Timer] | |
Setting range | 0.5 to 999 s |