| model (e.g. of a vehicle) |
SXN401-E |
SXN402-E |
SXN403-E |
SXN412-E |
SLN401-E |
SLN402-E |
SLN403-E |
SLN412-E |
| Temperature range |
-40C to +100C |
| Humidity range |
20%RH to 98%RH |
– |
| Temperature rise time |
Within 35 minutes-40C to +100C |
| Temperature fall time |
Within 55 minutes+20C to -40C |
Within 50 minutes+20C to -40C |
Within 55 minutes+20C to -40C |
Within 50 minutes+20C to -40C |
| Temperature change rate (rise) |
4.0C/min-26C to +86C |
| Temperature change rate (descending) |
2.5C/min+86C to -26C |
3.0C/min+86C to -26C |
2.5C/min+86C to -26C |
3.0C/min+86C to -26C |
| Temperature and humidity fluctuations |
}0.3C/}2.5%RH |
}0.3C |
| Temperature and humidity gradient |
3.0 in temperature operationb2.0/5.0%RH in temperature/humidity operation |
During temperature operation 3.0C |
| Space temperature and humidity deviation |
2.5 during temperature operationb1.5/5.0%RH during temperature/humidity operation |
During temperature operation 2.5C |
| Operable ambient temperature range |
+5C to +40C |
| Internal dimensions (W x H x D mm) |
500 x 350 x 350 |
(500 x 350 x 350) x 2 |
(500~350~350)~3 |
(600 x 600 x 500) x 2 |
500 x 350 x 350 |
(500 x 350 x 350) x 2 |
(500~350~350)~3 |
(600 x 600 x 500) x 2 |
| Content (liter) |
61 |
61 x 2 |
61 x 3 |
180 x 2 |
61 |
61 x 2 |
61 x 3 |
180 x 2 |
| External dimensions (W x H x D mm) |
660 x 1285 x 1040 |
660 x 1885 x 1040 |
660 x 1955 x 1375 |
760 x 1855 x 1645 |
660 x 1285 x 1040 |
660 x 1885 x 1040 |
660 x 1955 x 1375 |
760 x 1855 x 1645 |
| Mass (kg) |
190 |
330 |
500 |
475 |
185 |
325 |
495 |
470 |
| Exterior Steel Plates |
bonder steel sheet(Baked finish) |
| Inner tank steel plate |
Stainless steel plate(SUS304) |
| ventilator |
sirocco fan |
| refrigerator |
DC inverter controlled air-cooled rotary chiller |
| refrigerant |
R-448A |
| Heating Heater |
Nichrome Strip Heaters |
| humidifier |
sheathed heater(SUS316L) |
– |
| cooler |
Plate Fin Coil |
| Control Sensors |
JIS C1604 3-wire resistance thermometer Pt100 |
| Control method |
PID control |
| Display method |
LED indication |
| Display resolution |
0.1C/1%rh |
0.1C |
| Setting method |
Key-in method using the main unitfs display unit / input method using Smart Connect (optional) or a terminal equipped with a Web browser |
| Ethernet |
10BASE-T/100BASE-TX AUTO-MDIX compliant RJ45 connector x 1 channel |
| USB |
USB2.0 TYPE A connector x 1 channel |
| memory capacity |
8GB |
| power (button on TV, etc.) |
AC200V 3 50/60Hz Voltage fluctuation range }10% Class D grounding |
| humidified water |
Ion-exchanged water or pure water, etc.(0.1 to 10 S/cm) |
– |
| Maximum load current (A) |
15 |
30 |
45 |
46 |
9 |
18 |
27 |
32 |
| Operating current (A) |
11 |
22 |
33 |
34 |
7 |
14 |
21 |
26 |
| Exhaust heat (kW) |
0.62 to 3.6 |
1.24-7.2 |
1.86-10.8 |
1.0-9.6 |
0.62 to 3.6 |
1.24-7.2 |
1.86-10.8 |
1.0-9.6 |
| safeguard |
Main power leakage circuit breaker, fan motor overload relay, temperature overload protection device, refrigerator pressure switch, refrigerator overload relay, refrigerator overheat protection device, reverse phase detector, heating heater overload protection fuse, control circuit overload protection fuse, sensor disconnection detection function, temperature (humidity) upper/lower limit alarm function, power failure notification alarm, Filter cleaning alarm, Empty heating preventer, Humidifier heater overload protection fuse, Wick drying alarm, Water supply tank water level low alarm, Circuit protection ground-fault circuit breaker, Inverter abnormality detection function |
Main power leakage circuit breaker, fan motor overload relay, over-temperature protection device, refrigerator pressure switch, refrigerator overload relay, refrigerator overheat protection device, reverse phase detector, heating heater overload protection fuse, control circuit overload protection fuse, sensor disconnection detection function, upper and lower temperature limit alarm function, power failure notification alarm, filter cleaning alarm, circuit protection earth leakage circuit breaker, inverter error detection function |
| Ancillary Functions |
Memory backup function, pause function, monitor function, self-diagnostic function, power failure return protection function, instantaneous power failure retention function, external alarm output function, test termination output function, EUT power supply interlock function, auto-start/timer function, humidification delay function, standby operation function at room temperature |
Memory backup function, pause function, monitor function, self-diagnostic function, power failure return protection function, instantaneous power failure retention function, external alarm output function, test termination output function, EUT power supply interlock function, auto-start/timer function, standby operation function at room temperature |
| body (of a machine) |
Square measurement hole (100 x 30 mm, 2 locations on left and right sides), caster adjuster, dust filter for refrigerator condenser |
Square measurement hole (100 x 30 mm, 2 locations on left and right sides), caster adjuster, dust filter for refrigerator condenser |
| accessory |
Wick gauze (12 wicks/tank), silicon stopper for square measurement hole (2 stoppers/tank), adjuster fixture (4 fixtures/unit, depending on model), key for door latch (2 keys/tank), instruction manual (1 copy), water tank (401, 402: 10L, 403, 412: 20L, all models 1 tank/unit), certificate of inspection acceptance |
Silicone plugs for square measurement holes (2/tank), adjuster fixture (4/unit, depending on model), keys for door latches (2/tank), instruction manual (1 copy), inspection certificate |