This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.
ITEMLIST
Products Infomation
Capable of high-speed response pulse source, 5½-digit measurement and highprecision low-resistance measurement
The 6240B is a SMU equipped with a high-speed response pulse source and measurement function and a low-resistance measurement function. It has 4½-digit display for voltage and current source and 5½-digit display for measurement, and features high accuracy of ±0.02 %. There are four types of sweep modes available: liner, fixed, random and 2-slope linear sweep, and also pulse measurement with a minimum pulse width of 50 µs is available. Thus, this model can be widely used as evaluation power supply for developing semiconductors and other electronic components and as power supply of characteristic test systems used in production lines.
Moreover, the 6240B is capable of high-precision contact resistance measurement that cancels thermal EMF generated on metal contact surfaces.
DC Voltage Current Source / Monitor
Model:
6240B
Maker:
ADC CORPORATION
Features
- Source and measurement range
- Voltage: 0 to ±15 V, Current: 0 to ±4 A (1 A for DC)
- Source and measurement accuracy: ±0.02 %
- Measurement display of 5½ digits with resolution of 1 μV/100 pA
- Source/sink of ±4 A for the maximum pulse width of 20 ms
- Pulse measurement with minimum pulse width of 50 μs and 1 μs step
- Sink-enabled bipolar output
- Low-resistance measurement canceling thermal electromotive
force for conductive materials
Specifications
| Number of digits for generation | 4½ | ||
| Output method | Bipolar | ||
| Maximum output (top) | Voltage | ±15 V/1 A 10 μV |
|
| Minimum resolution (bottom) | Current | ±4 A/10 V (DC: ±1 A/15 V) 1 nA |
|
| Number of digits for measurement | 5½ | ||
| Accuracy (typical range) | 0.02% | ||
| Minimum measurement resolution | Voltage | 1 μV | |
| Current | 100 pA | ||
| Maximum measurement resistnace/ Minimum resolution | 0.75 GΩ/0.25 μΩ | ||
| Pulse application / Measurement | Available | ||
| Minimum pulse width | 50 μs | ||
| Interface | USB, GPIB | ||